Determination of airborne carbonyls: Comparison of a thermal desorption/GC method with the standard DNPH/HPLC method

被引:122
作者
Ho, SSH [1 ]
Yu, JZ [1 ]
机构
[1] Hong Kong Univ Sci & Technol, Dept Chem, Kowloon, Hong Kong, Peoples R China
关键词
D O I
10.1021/es034795w
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The standard method for the determination of gaseous carbonyls is to collect carbonyls onto 2,4-dinitrophenyl hydrazine (DNPH) coated solid sorbent followed by solvent extraction of the solid sorbent and analysis of the derivatives using high-pressure liquid chromatography (HPLC). This paper describes a newly developed approach that involves collection of the carbonyls onto pentafluorophenyl hydrazine (PFPH) coated solid sorbents followed by thermal desorption and gas chromatographic (GC) analysis of the PFPH derivatives with mass spectrometric (MS) detection. Sampling tubes loaded with 510 nmol of PFPH on Tenax sorbent effectively collect gaseous carbonyls, including formaldehyde, acetaldehyde, propanal, butanal, heptanal, octanal, acrolein, 2-furfural, benzaldehyde, p-tolualdehyde, glyoxal, and methylglyoxal, at a flow rate of at least up to 100 mL/min. All of the tested carbonyls are shown to have method detection limits (MDLs) of sub-nanomoles per sampling tube, corresponding to air concentrations of < 0.3 ppbv for a sampled volume of 24 L. These limits are 2-12 times lower than those that can be obtained using the DNPH/HPLC method. The improvement of MDLs is especially pronounced for carbonyls larger than formaldehyde and acetaldehyde. The PFPH/GC method also offers better peak separation and more sensitive and specific detection through the use of MS detection. Comparison studies on ambient samples and kitchen exhaust samples have demonstrated that the two methods do not yield systematic differences in concentrations of the carbonyls that are above their respective MDLs in both methods, including formaldehyde, acetaldehyde, acrolein, and butanal. The lower MDLs afforded by the PFPH/GC method also enable the determination of a few more carbonyls in both applications.
引用
收藏
页码:862 / 870
页数:9
相关论文
共 27 条
[1]   Observation of volatile and semi-volatile carbonyls in an Algerian urban environment using dinitrophenylhydrazine/silica-HPLC and pentafluorophenylhydrazine/silica-GC-MS [J].
Cecinato, A ;
Yassaa, N ;
Di Palo, V ;
Possanzini, M .
JOURNAL OF ENVIRONMENTAL MONITORING, 2002, 4 (02) :223-228
[2]   Pentafluorophenylhydrazine as a coating reagent for the HRGC-MS determination of semi volatile carbonyl compounds in air [J].
Cecinato, A ;
Di Palo, V ;
Mabilia, R ;
Possanzini, M .
CHROMATOGRAPHIA, 2001, 54 (3-4) :263-269
[3]  
DYE C, 1992, HRC-J HIGH RES CHROM, V15, P5
[4]   DETERMINATION OF NANOGRAM AMOUNTS OF CARBONYLS AS 2,4-DINITROPHENYLHYDRAZONES BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
FUNG, K ;
GROSJEAN, D .
ANALYTICAL CHEMISTRY, 1981, 53 (02) :168-171
[5]   COLLECTION EFFICIENCIES OF CARTRIDGES AND MICRO-IMPINGERS FOR SAMPLING OF ALDEHYDES IN AIR AS 2,4-DINITROPHENYLHYDRAZONES [J].
GROSJEAN, D ;
FUNG, K .
ANALYTICAL CHEMISTRY, 1982, 54 (07) :1221-1224
[6]   AMBIENT LEVELS OF FORMALDEHYDE, ACETALDEHYDE, AND FORMIC-ACID IN SOUTHERN CALIFORNIA - RESULTS OF A ONE-YEAR BASE-LINE STUDY [J].
GROSJEAN, D .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1991, 25 (04) :710-715
[7]   Air quality model evaluation data for organics .2. C-1-C-14 carbonyls in Los Angeles air [J].
Grosjean, E ;
Grosjean, D ;
Fraser, MP ;
Cass, GR .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1996, 30 (09) :2687-2703
[8]   Carbonyl collection efficiency of the DNPH-coated C-18 cartridge in dry air and in humid air [J].
Grosjean, E ;
Grosjean, D .
ENVIRONMENTAL SCIENCE & TECHNOLOGY, 1996, 30 (03) :859-863
[9]   Concentrations of formaldehyde and other carbonyls in environments affected by incense burning [J].
Ho, SSH ;
Yu, JZ .
JOURNAL OF ENVIRONMENTAL MONITORING, 2002, 4 (05) :728-733
[10]   Feasibility of collection and analysis of airborne carbonyls by on-sorbent derivatization and thermal desorption [J].
Ho, SSH ;
Yu, JZ .
ANALYTICAL CHEMISTRY, 2002, 74 (06) :1232-1240