Assessment of coronary arterial flow and flow reserve in humans with magnetic resonance imaging

被引:135
作者
Hundley, WG
Lange, RA
Clarke, GD
Meshack, BM
Payne, J
Landau, C
McColl, R
Sayad, DE
Willett, DL
Willard, JE
Hillis, LD
Peshock, RM
机构
[1] UNIV TEXAS, SW MED CTR, MARY NELL & RALPH B ROGERS MAGNET RESONANCE CTR, DEPT INTERNAL MED, DALLAS, TX 75235 USA
[2] UNIV TEXAS, SW MED CTR, DEPT RADIOL, DALLAS, TX 75235 USA
关键词
magnetic resonance imaging; coronary disease; regional blood flow;
D O I
10.1161/01.CIR.93.8.1502
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background The noninvasive measurement of absolute epicardial coronary arterial flow and flow reserve would be useful in the evaluation of patients with coronary circulatory disorders. Phase-contrast magnetic resonance imaging (PC-MRI) has been used to measure coronary arterial flow in animals, but its accuracy in humans is unknown. Methods and Results Twelve subjects (7 men: 5 women; age, 44 to 67 years) underwent PC-MRI measurements of flow in the left anterior descending coronary artery or one of its diagonal branches at rest and after administration of adenosine (140 mu g . kg(-1). min(-1) IV). Immediately thereafter, intracoronary Doppler velocity (IDV) and flow measurements were made during cardiac catheterization at rest and after intravenous administration of adenosine. For the 12 patients, the correlation between MRI and invasive measurements of coronary arterial flow and coronary arterial flow reserve was excellent: coronary flow(MRI) (mL/min)=0.85 X coronary flow(IDV) (mL/min)+17 (mL/min), r=.89, and coronary flow reserve(MRI)=0.79 X coronary velocity reserve(IDV)+0.34, r=.89. For the range of coronary arterial flows (18 to 161 mL/min) measured by MRI, the limit of agreement between MRI and catheterization measurements of flow was -13+/-30 mL/min; for the range of coronary reserves (0.7 to 3.7) measured by MRI, the limit of agreement between the two techniques was 0.1+/-0.4. Conclusions Cine velocity-encoded PC-MRI can noninvasively measure absolute coronary arterial flow in the left anterior descending artery in humans. PC-MRI can detect pharmacologically induced changes in coronary arterial flow and can reliably distinguish between those subjects with normal and abnormal coronary artery flow reserve.
引用
收藏
页码:1502 / 1508
页数:7
相关论文
共 37 条
[1]  
ARLTIER DC, 1981, AM J PHYSIOL, V240, pH767
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   MYOCARDIAL STUNNING IN MAN [J].
BOLLI, R .
CIRCULATION, 1992, 86 (06) :1671-1691
[4]   MEASUREMENT OF ABSOLUTE EPICARDIAL CORONARY-ARTERY FLOW AND FLOW RESERVE WITH BREATH-HOLD CINE PHASE-CONTRAST MAGNETIC-RESONANCE-IMAGING [J].
CLARKE, GD ;
ECKELS, R ;
CHANEY, C ;
SMITH, D ;
DITTRICH, J ;
HUNDLEY, WG ;
NESSAIVER, M ;
LI, HF ;
PARKEY, RW ;
PESHOCK, RM .
CIRCULATION, 1995, 91 (10) :2627-2634
[5]   VALIDATION OF A DOPPLER GUIDE WIRE FOR INTRAVASCULAR MEASUREMENT OF CORONARY-ARTERY FLOW VELOCITY [J].
DOUCETTE, JW ;
CORL, PD ;
PAYNE, HM ;
FLYNN, AE ;
GOTO, M ;
NASSI, M ;
SEGAL, J .
CIRCULATION, 1992, 85 (05) :1899-1911
[6]   FLOW VELOCITY QUANTIFICATION IN HUMAN CORONARY-ARTERIES WITH FAST, BREATH-HOLD MR-ANGIOGRAPHY [J].
EDELMAN, RR ;
MANNING, WJ ;
GERVINO, E ;
LI, W .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1993, 3 (05) :699-703
[7]   CORONARY-ARTERIES - BREATH-HOLD MR ANGIOGRAPHY [J].
EDELMAN, RR ;
MANNING, WJ ;
BURSTEIN, D ;
PAULIN, S .
RADIOLOGY, 1991, 181 (03) :641-643
[8]   AN IMPROVED QUADRATURE OR PHASED-ARRAY COIL FOR MR CARDIAC IMAGING [J].
FAYAD, ZA ;
CONNICK, TJ ;
AXEL, L .
MAGNETIC RESONANCE IN MEDICINE, 1995, 34 (02) :186-193
[9]   INVIVO VALIDATION OF MR VELOCITY IMAGING [J].
FIRMIN, DN ;
NAYLER, GL ;
KLIPSTEIN, RH ;
UNDERWOOD, SR ;
REES, RSO ;
LONGMORE, DB .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1987, 11 (05) :751-756
[10]  
GREENBERG MA, 1981, CIRCULATION, V64, P305