Design and characterization of a dynamic multileaf collimator

被引:10
作者
Loi, G
Pignoli, E
Scorsetti, M
Cerreta, V
Somigliana, A
Marchesini, R
Gramaglia, A
Cerchiari, U
Ricci, SB
机构
[1] Ist Nazl Studio & Cura Tumori, Div Hlth Phys, I-20133 Milan, Italy
[2] Ist Nazl Studio & Cura Tumori, Div Radiotherapy B, I-20133 Milan, Italy
关键词
D O I
10.1088/0031-9155/43/10/033
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The characteristics of a prototype computer-assisted dynamic multileaf collimator (DMLC), specifically designed for small-held conformal radiotherapy, were evaluated at the Istituto Nazionale Tumori of Milan. The collimating device consists of two opposing banks of 16 pairs of 8 cm thick, 3.6 mm wide tungsten leaves and allows shaping of a radiation field up to a size of 10 x 10 cm(2) at the isocentre. The screening thickness of each leaf is 6.25 mm at the accelerator gantry isocentre. The leaves have a trapezoidal cross section and move along an arched path, thus providing a 'double focused' collimation system. The DMLC was installed on the head of a Varian Clinac 2100C linear accelerator. Mechanical and dosimetric evaluations were performed to test the stability of the mechanical isocentre and to determine leaf leakage, penumbra width, accuracy of leaf positions and uniformity of leaf speed. Displacement of the mechanical isocentre was less than 1 mm at all gantry angles. Standard radiographic films exposed to 6 MV x-ray radiation were used for dosimetric evaluations. Leakage between leaves was less than 2.5%, and leakage through abutted leaves was less than 5.5%. The penumbra width between 20% and 80% isodose at different positions of leaf banks was 2.7 mm in the direction of the leaf motion and 3.1 mm along the side of the leaf with a standard deviation of 0.2 mm in both directions. Accuracy in the positioning of the leaf was 0.3 mm, whereas the maximum repositioning error was less than 0.2 mm. Finally, during movement of the leaves at the maximum speed of 0.5 mm s(-1), the standard deviation of the leaf positioning error was 0.2 mm, proving an accurate uniformity of leaf speed.
引用
收藏
页码:3149 / 3155
页数:7
相关论文
共 13 条
[1]   COMPARISON OF THE PENUMBRA BETWEEN FOCUSED AND NONDIVERGENT BLOCKS - IMPLICATIONS FOR MULTILEAF COLLIMATORS [J].
BIGGS, P ;
CAPALUCCI, J ;
RUSSELL, M .
MEDICAL PHYSICS, 1991, 18 (04) :753-758
[2]   STATIC-FIELD CONFORMAL STEREOTAXIC RADIOSURGERY - PHYSICAL TECHNIQUES [J].
BOURLAND, JD ;
MCCOLLOUGH, KP .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (02) :471-479
[3]   EVALUATION OF MULTILEAF COLLIMATOR DESIGN FOR A PHOTON-BEAM [J].
GALVIN, JM ;
SMITH, AR ;
MOELLER, RD ;
GOODMAN, RL ;
POWLIS, WD ;
RUBENSTEIN, J ;
SOLIN, LJ ;
MICHAEL, B ;
NEEDHAM, M ;
HUNTZINGER, CJ ;
KLIGERMAN, MM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1992, 23 (04) :789-801
[4]   CHARACTERIZATION OF A MULTILEAF COLLIMATOR SYSTEM [J].
GALVIN, JM ;
SMITH, AR ;
LALLY, B .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (02) :181-192
[5]   Dosimetric comparison of an integrated multileaf-collimator versus a conventional collimator [J].
Georg, D ;
Julia, F ;
Briot, E ;
Huyskens, D ;
Wolff, U ;
Dutreix, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (11) :2285-2303
[6]   THE DESIGN AND PERFORMANCE-CHARACTERISTICS OF A MULTILEAF COLLIMATOR [J].
JORDAN, TJ ;
WILLIAMS, PC .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (02) :231-251
[7]   Clinical implementation of a commercial multileaf collimator: Dosimetry, networking, simulation, and quality assurance [J].
Klein, EE ;
Harms, WB ;
Low, DA ;
Willcut, V ;
Purdy, JA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (05) :1195-1208
[8]  
MALEKI N, 1984, MED PHYS, V11, P390
[9]   VARIABLES ASSOCIATED WITH THE DEVELOPMENT OF COMPLICATIONS FROM RADIOSURGERY OF INTRACRANIAL TUMORS [J].
NEDZI, LA ;
KOOY, H ;
ALEXANDER, E ;
GELMAN, RS ;
LOEFFLER, JS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (03) :591-599
[10]   COMPUTER-SYSTEMS AND MECHANICAL TOOLS FOR STEREOTAXICALLY GUIDED CONFORMATION THERAPY WITH LINEAR ACCELERATORS [J].
SCHLEGEL, W ;
PASTYR, O ;
BORTFELD, T ;
BECKER, G ;
SCHAD, L ;
GADEMANN, G ;
LORENZ, WJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1992, 24 (04) :781-787