Late dissociation of the polyethylene liner from a modular acetabular metal shell after primary total hip arthroplasty - A report of five cases

被引:23
作者
Han, CD [1 ]
Choe, WS [1 ]
Yoo, JH [1 ]
机构
[1] Yonsei Univ, Coll Med, Dept Orthopaed Surg, Seoul 120752, South Korea
关键词
total hip replacement; modular prosthesis; polyethylene dissociation; acetabular metal shell; locking mechanism;
D O I
10.3349/ymj.1998.39.3.277
中图分类号
R5 [内科学];
学科分类号
1002 [临床医学]; 100201 [内科学];
摘要
Modular designs of hip prostheses have become popular recently. Along with complications inherent in all hip arthroplasty systems, modular systems have the additional potential for dissociation of components. Five male patients underwent total hip arthroplasties, in which all of the acetabular components were Harris-Galante II porous acetabular cups. Many years after the operation, the polyethylene liners were dissociated without any previous trauma or dislocation of the femoral heads. These dissociations and dislodgements were managed with open reduction. This complication can be predicted from clinical symptoms and signs. Roentgenograms must be taken and carefully compared to previous roentgenograms. We postulated two causes for the dissociation. First, the polyethylene liner war not fixed securely within the acetabular metal shell at the time of operation. Second, the locking mechanism of the acetabular metal shell was not strong enough to firmly hold the polyethylene liner within the acetabular metal shell. It does warrant that certain precautions must be taken when implanting modular components. The locking mechanism of the Harris-Galante II porous acetabular component is mechanically weak and fails easily, therefore its design must be improved in an attempt to prevent postoperative dissociation of the polyethylene liner.
引用
收藏
页码:277 / 282
页数:6
相关论文
共 12 条
[1]
DISLOCATION AND COMPONENT SEPARATION OF BATEMAN HIP ENDOPROSTHESIS [J].
ANDERSON, PR ;
MILGRAM, JW .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1978, 240 (19) :2079-2080
[2]
BARMADA R, 1987, CLIN ORTHOP RELAT R, P147
[3]
DISASSEMBLY OF A ONE-PIECE METAL-BACKED ACETABULAR COMPONENT - A CASE-REPORT [J].
BEAVER, RJ ;
SCHEMITSCH, EH ;
GROSS, AE .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1991, 73 (06) :908-910
[4]
DISSOCIATION OF ACETABULAR COMPONENTS AFTER TOTAL HIP-ARTHROPLASTY - REPORT OF 4 CASES [J].
BRIEN, WW ;
SALVATI, EA ;
WRIGHT, TM ;
NELSON, CL ;
HUNGERFORD, DS ;
GILLIAM, DL .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (10) :1548-1550
[5]
Bueche M J, 1989, J Arthroplasty, V4, P39, DOI 10.1016/S0883-5403(89)80052-X
[6]
Cameron H U, 1993, Orthop Rev, V22, P1160
[7]
Fisher D A, 1994, J Arthroplasty, V9, P325, DOI 10.1016/0883-5403(94)90088-4
[8]
DISASSEMBLY OF A MODULAR ACETABULAR COMPONENT OF A TOTAL HIP-REPLACEMENT ARTHROPLASTY - A CASE-REPORT [J].
KITZIGER, KJ ;
DELEE, JC ;
EVANS, JA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (04) :621-623
[9]
MLLER BN, 1983, ACTA ORTHOP SCAN, V54, P553
[10]
DISSOCIATION OF MODULAR HIP-ARTHROPLASTY COMPONENTS AFTER DISLOCATION - A REPORT OF 3 CASES AT DIFFERING DISSOCIATION LEVELS [J].
STAR, MJ ;
COLWELL, CW ;
DONALDSON, WF ;
WALKER, RH .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1992, (278) :111-115