Helical tomotherapy radiation leakage and shielding considerations

被引:28
作者
Balog, J
Lucas, D
DeSouza, C
Crilly, R
机构
[1] TomoTherapy Inc, Madison, WI 53717 USA
[2] Univ Wisconsin, Dept Med Phys, Madison, WI 53706 USA
[3] Rapid City Reg Hosp, Rapid City, SD 57701 USA
关键词
D O I
10.1118/1.1861521
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Leakage radiation and room shielding considerations increase significantly for intensity-modulated radiation therapy (IMRT) treatments due to the increased beam-on time to deliver modulated fields. Tomotherapy, with its slice by slice approach to IMRT, further exacerbates this increase. Accordingly, additional shielding is used in tomotherapy machines to reduce unwanted radiation. The competing effects of the high modulation and the enhanced shielding were studied. The overall room leakage radiation levels are presented for the continuous gantry rotations, which are always used during treatments. The measured leakage at 4 m from the isocenter is less than 3 x 10(-4) relative to calibration output. Primary radiation exposure levels were investigated as well. The effect of forward-directed leakage through the beam-collimation system was studied, as this is the leakage dose the patient would receive in the course of a treatment. A 12-min treatment was calculated to produce only 1% patient leakage dose to the periphery region. Longer treatment times might yield less patient dose if the field width selected is correspondingly narrower. A method for estimating the worst-case leakage dose a patient would receive is presented. (c) 2005 American Association of Physicists in Medicine.
引用
收藏
页码:710 / 719
页数:10
相关论文
共 16 条
[1]  
[Anonymous], 1993, 116 NCRP MEAS
[2]   Benchmarking beam alignment for a clinical helical tomotherapy device [J].
Balog, J ;
Mackie, TR ;
Pearson, D ;
Hui, S ;
Paliwal, B ;
Jeraj, R .
MEDICAL PHYSICS, 2003, 30 (06) :1118-1127
[3]   Estimates of whole-body dose equivalent produced by beam intensity modulated conformal therapy [J].
Followill, D ;
Geis, P ;
Boyer, A .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (03) :667-672
[4]   Optimization of helical tomotherapy treatment plans for prostate cancer [J].
Grigorov, G ;
Kron, T ;
Wong, E ;
Chen, J ;
Sollazzo, J ;
Rodrigues, G .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (13) :1933-1943
[5]   Radiation-induced second cancers: The impact of 3D-CRT and IMRT [J].
Hall, EJ ;
Wuu, CS .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 56 (01) :83-88
[6]  
*INT EL COMM, 1987, IEC PUBL, V60
[7]   Tomotherapy [J].
Mackie, TR ;
Balog, J ;
Ruchala, K ;
Shepard, D ;
Aldridge, S ;
Fitchard, E ;
Reckwerdt, P ;
Olivera, G ;
McNutt, T ;
Mehta, M .
SEMINARS IN RADIATION ONCOLOGY, 1999, 9 (01) :108-117
[8]   TOMOTHERAPY - A NEW CONCEPT FOR THE DELIVERY OF DYNAMIC CONFORMAL RADIOTHERAPY [J].
MACKIE, TR ;
HOLMES, T ;
SWERDLOFF, S ;
RECKWERDT, P ;
DEASY, JO ;
YANG, J ;
PALIWAL, B ;
KINSELLA, T .
MEDICAL PHYSICS, 1993, 20 (06) :1709-1719
[9]  
MCGINLEY PH, 1998, SHIELDING TECHNIQUES, P18
[10]   Whole-body dose from tomotherapy delivery [J].
Mutic, S ;
Low, DA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 42 (01) :229-232