Whole-body dose from tomotherapy delivery

被引:72
作者
Mutic, S [1 ]
Low, DA [1 ]
机构
[1] Edward Mallinckrodt Inst Radiol, Div Radiat Oncol, St Louis, MO 63110 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1998年 / 42卷 / 01期
关键词
intensity modulated radiation therapy; whole-body dose; tomotherapy;
D O I
10.1016/S0360-3016(98)00199-0
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To measure whole-body dose in tomotherapy of the head and neck region resulting from internal patient scatter and linear accelerator leakage. Methods and Materials: Treatments are performed using a commercial computer-controlled intensity modulated radiation therapy planning and delivery system (Peacock, NOMOS Corp.) and a 6-MV linear accelerator (Clinac 6/100, Varian Corp.). The patient dose outside the treatment held is measured in a water-equivalent phantom using thermoluminescent dosimetry. The whole-body dose components from internal scatter and leakage are separately determined. The use of fixed-portal leakage and scattered radiation measurements to estimate the whale-body dose from tomotherapy is evaluated. Results: The internally scattered dose is significant near the target, but becomes negligible relative to the leakage dose beyond 15 cm from the target. Dose at 10 cm from the target volume, due to internal scatter and leakage, is approximately 2.5% of the total target dose, reducing to 0.5% at 30 cm. The measured dose is relatively uniform throughout the phantom, Conclusion: The whole-body dose equivalent from a tomotherapy treatment is greater than that from conventional. radiation therapy. Further studies are required to assess the trade-off between improved dose distribution conformality and a possible slight increase in radiation-induced fatal malignancies. The accuracy of using fixed-portal leakage and scattered dose measurements to estimate the whole-body dose from tomotherapy treatments is adequate, if the appropriate fixed-portal field size equivalent is used. (C) 1998 Elsevier Science Inc.
引用
收藏
页码:229 / 232
页数:4
相关论文
共 10 条
[1]  
BUTLER EB, 1994, RADIOLOGY, V193, P228
[2]   The field-matching problem as it applies to the peacock three dimensional conformal system for intensity modulation [J].
Carol, M ;
Grant, WH ;
Bleier, AR ;
Kania, AA ;
Targovnik, HS ;
Butler, EB ;
Woo, SW .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 34 (01) :183-187
[3]  
CAROL MP, 1993, 3D RAD TREATMENT PLA
[4]   Estimates of whole-body dose equivalent produced by beam intensity modulated conformal therapy [J].
Followill, D ;
Geis, P ;
Boyer, A .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (03) :667-672
[5]  
Followill DS, 1997, INT J RADIAT ONCOL, V39, P783
[6]  
GRANT W, 1996, P 1996 AAPM SUMM SCH
[7]   Phantoms for IMRT dose distribution measurement and treatment verification [J].
Low, DA ;
Gerber, RL ;
Mutic, S ;
Purdy, JA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 40 (05) :1231-1235
[8]   FETAL DOSE FROM RADIOTHERAPY WITH PHOTON BEAMS - REPORT OF AAPM-RADIATION-THERAPY-COMMITTEE TASK GROUP NO-36 [J].
STOVALL, M ;
BALCKWELL, CR ;
CUNDIFF, J ;
NOVACK, DH ;
PALTA, JR ;
WAGNER, LK ;
WEBSTER, EW ;
SHALEK, RJ .
MEDICAL PHYSICS, 1995, 22 (01) :63-82
[9]   Initial experience with intensity-modulated conformal radiation therapy for treatment of the head and neck region [J].
Verellen, D ;
Linthout, N ;
VandenBerge, D ;
Bel, A ;
Storme, G .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 39 (01) :99-114
[10]   Evaluation of dose calculation algorithm of the peacock system for multileaf intensity modulation collimator [J].
Wu, A ;
Johnson, M ;
Chen, ASJ ;
Kalnicki, S .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 36 (05) :1225-1231