Feasibility of an on-line restricted access material/liquid chromatography/tandem mass spectrometry method in the rapid and sensitive determination of organophosphorus triesters in human blood plasma

被引:45
作者
Amini, N [1 ]
Crescenzi, C [1 ]
机构
[1] Univ Stockholm, Dept Analyt Chem, S-10691 Stockholm, Sweden
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 2003年 / 795卷 / 02期
关键词
restricted access material; organophosphorus triesters;
D O I
10.1016/S1570-0232(03)00568-3
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A rapid on-line solid phase extraction/liquid chroinatography/tandem mass spectrometry (SPE/LC/MS/MS) method using restricted access material (RAM) was developed for the simultaneous determination of eight organophosphorus triesters in untreated human blood plasma. In a process involving column-switching techniques, the analytes were enriched on the RAM column, separated using a C-18 analytical column and detected with LC/MS. Tandem mass spectrometry was used to characterize and quantify the analytes. To elucidate the fragmentation pathway of a number of the analytes, MS3 experiments using an ion trap mass spectrometer were performed. The matrix effects associated with using APCI and ESI interfaces were investigated. The recoveries obtained were in the range 60-92% (R.S.D. < 6%), with estimated detection limits between 0.2 and 1.8 ng/ml of plasma, and the total analysis time was 27 min. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:245 / 256
页数:12
相关论文
共 39 条