DMX: An absolutely calibrated time-resolved broadband soft x-ray spectrometer designed for MJ class laser-produced plasmas (invited)

被引:95
作者
Bourgade, JL
Villette, B
Bocher, JL
Boutin, JY
Chiche, S
Dague, N
Gontier, D
Jadaud, JP
Savale, B
Wrobel, R
Turner, RE
机构
[1] CEA, DAM Ile France, F-92160 Bruyeres Le Chatel, France
[2] Lawrence Livermore Natl Lab, Livermore, CA 94550 USA
关键词
D O I
10.1063/1.1324744
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In high flux (10(13-15) W/cm(2)) laser-plasma interaction, a large part of the incoming laser energy is radiated into soft x-rays. To determine both the shape of this spectrum and the conversion efficiency, we have designed and tested an absolutely calibrated broadband soft x-ray spectrometer with a high temporal resolution (100 ps). The detector in this spectrometer is a new coaxial x-ray diode coupled with a fast single shot oscilloscope (French IN7100 with 7 GHz frequency response cutoff). Both absolute calibrations (x-ray response of diodes) and relative calibrations (filters and mirrors) have used the French synchrotron beam lines at Laboratoire pour l'utilisation du Rayonnement Electromagnetique (LURE) in Orsay. The initial version of this instrument was first successfully implemented on laser plasmas experiments at the Phebus facility in France and an improved version is now operating at the Omega laser facility in Rochester, New York. The emitted x-ray spectrum is absolutely measured in 18 broad bands from 50 eV up to 20 keV. The softer bands (<1.5 keV) combine mirror and filter responses coupled with the coaxial diode response to improve hard x-ray rejection. Intermediate energy channels (1.5 keV <h nu <5 keV) used only a filter and coaxial diode. For the hardest channels (>5 keV) we replace the x-ray diode (not sufficiently sensitive) with a photoconductive detector (neutron-damaged GaAs). An equivalent instrument will be designed for the future National Ignition Facility (NIF) in the United States and the Laser Megajoule (LMJ) in France. (C) 2001 American Institute of Physics.
引用
收藏
页码:1173 / 1182
页数:10
相关论文
共 35 条
[1]  
ALHSTROM HG, 1979, J OPT SOC AM, V68, P1731
[2]   X-ray conversion with PHEBUS laser [J].
Babonneau, D ;
Bonnet, L ;
Jacquemot, S ;
Bocher, JL ;
Boutin, JY ;
Jadaud, JP ;
Vilette, B .
LASER AND PARTICLE BEAMS, 1999, 17 (03) :459-463
[3]  
BABONNEAU D, 1984, LASER INTERACTION RE, V6, P817
[4]   X-ray detector calibration at CEA/Bruyeres-le-Chatel [J].
Beck, L ;
Stemmler, P ;
Ban, G ;
Villette, B ;
Frotte, V ;
Bizeuil, C ;
Boutin, JY ;
Nazet, C .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 369 (2-3) :401-406
[5]   SUPER-ACO - RESULTS ON A POSITRON LOW EMITTANCE RING [J].
BESSON, JC ;
CERTAIN, P ;
DAEL, A ;
DAMANY, A ;
JUAN, P ;
LABEQUE, A ;
LEVEL, MP ;
MARIN, PC ;
MICHAUT, J ;
MONETDESCOMBEY, C ;
NGHIEM, P ;
SOMMER, M ;
SOUCHET, R ;
ZYNGIER, H .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (07) :1373-1376
[6]  
BOCHER JL, 1980, RAPPORT ACTIVITES LA, P86
[7]  
BOCHER JL, 1999, REV SCI TECHNIQUE DI, P11
[8]  
BOURGADE JL, 1985, CR ACAD SCI II, V301, P895
[9]  
BOURGADE JL, 1998, Patent No. 9843275
[10]   Discharge from hot CaO. [J].
Child, CD .