Effects of Steris 1™ sterilization and Cidex® ortho-phthalaldehyde high-level disinfection on durability of new-generation flexible ureteroscopes

被引:27
作者
Abraham, Jose Benito A.
Abdelshehid, Corollos S.
Lee, Hak J.
Box, Geoffrey N.
Deane, Leslie A.
Le, Todd
Jellison, Forrest
Borin, James F.
Manipon, Anthony
McDougall, Elspeth M.
Clayman, Ralph V.
机构
[1] Univ Calif Irvine, Dept Urol, Orange, CA 92868 USA
[2] Loma Linda Univ, Sch Med, Loma Linda, CA 92350 USA
[3] Univ Maryland, Dept Urol, Baltimore, MD 21201 USA
关键词
D O I
10.1089/end.2007.0181
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 [临床医学]; 100201 [内科学];
摘要
Background and Purpose: The effects of commonly used reprocessing methods on flexible ureteroscope longevity have never been examined. We prospectively studied the effects of Steris 1(TM) sterilization and Cidex(R) ortho-phthalaldehyde (OPA) high-level disinfection (HLD) on the image quality, physical structure, and deflective properties of two new flexible ureteroscopes. Materials and Methods: Two identical "out-of-the-box" Storz 11278AU1 flexible ureteroscopes (Karl Storz Endoscopy, Tuttlingen, Germany) were sterilized individually using the Steris 1(TM) system (Steris Mentor, Ohio) or disinfected with Cidex(R) OPA (Advanced Sterilization Products, J&J, Irvine, CA) for 100 trials followed by a crossover to the other method for another 100 trials over a period of 1 year. After every five trials, optical quality, angle of deflection, and fiber damage were analyzed in the laboratory. Throughout the study, neither of these ureteroscopes was used clinically. Results: After 100 trials, ureteroscope 1, which was sterilized initially in the Steris system, had a 12-mm tear on its shaft (noted after the 17(th) trial), 297 damaged fibers, and a 37% drop in resolution (loss of 3.75 lines/mm). There was no change in deflection from baseline. In contrast, after 100 cycles, ureteroscope 2, which was subjected to HLD with Cidex(R) OPA, had no visible external damage, a 0% change in resolution, 10 damaged fibers, and no change in deflection. After the crossover, ureteroscope 2 developed a semilunar defect that obscured the endoscopic view, whereas there was no further significant damage to ureteroscope 1. Conclusion: After 100 cycles, the Steris 1 system rendered the flexible ureteroscope unusable, whereas HLD with Cidex(R) OPA had minimal adverse impact.
引用
收藏
页码:985 / 992
页数:8
相关论文
共 28 条
[1]
Comparison of flexible ureteroscopes: Deflection, irrigant flow and optical characteristics [J].
Abdelshehid, C ;
Ahlering, MT ;
Chou, D ;
Park, HK ;
Basillote, J ;
Lee, D ;
Kim, I ;
Eichel, L ;
Protsenko, D ;
Wong, B ;
McDougall, E ;
Clayman, RV .
JOURNAL OF UROLOGY, 2005, 173 (06) :2017-2021
[2]
Flexible ureteroscopes: A single center evaluation of the durability and function of the new endoscopes smaller than 9Fr [J].
Afane, JS ;
Olweny, EO ;
Bercowsky, E ;
Sundaram, CP ;
Dunn, MD ;
Shalhav, AL ;
McDougall, EM ;
Clayman, RV .
JOURNAL OF UROLOGY, 2000, 164 (04) :1164-1168
[3]
Importance of lumen flow in liquid chemical sterilization [J].
Alfa, MJ .
AMERICAN JOURNAL OF INFECTION CONTROL, 1999, 27 (04) :373-374
[4]
IN-HOSPITAL EVALUATION OF ORTHOPHTHALALDEHYDE AS A HIGH-LEVEL DISINFECTANT FOR FLEXIBLE ENDOSCOPES [J].
ALFA, MJ ;
SITTER, DL .
JOURNAL OF HOSPITAL INFECTION, 1994, 26 (01) :15-26
[5]
Brown Stanley A., 2002, Biomedical Instrumentation and Technology, V36, P23, DOI 10.2345/0899-8205(2002)36[23:EODDAS]2.0.CO
[6]
2
[7]
Frequency of Ureteroscope damage seen at a tertiary care center [J].
Carey, Robert I. ;
Gomez, Christopher S. ;
Maurici, Giuseppe ;
Lynne, Charles M. ;
Leveillee, Raymond J. ;
Bird, Vincent G. .
JOURNAL OF UROLOGY, 2006, 176 (02) :607-610
[8]
An evaluation of Cidex OPA (0.55% ortho-phthalaldehyde) as an alternative to 2% glutaraldehyde for high-level disinfection of endoscopes [J].
Cooke, RPD ;
Goddard, SV ;
Whymant-Morris, A ;
Sherwood, J ;
Chatterly, R .
JOURNAL OF HOSPITAL INFECTION, 2003, 54 (03) :226-231
[9]
Evaluation of the microbicidal efficacy of STERIS SYSTEM I for digestive endoscopes using GERMANDE and ASTM validation protocols [J].
Duc, DL ;
Ribiollet, A ;
Dode, X ;
Ducel, G ;
Marchetti, B ;
Calop, J .
JOURNAL OF HOSPITAL INFECTION, 2001, 48 (02) :135-141
[10]
Efficacy of high-level disinfectants for reprocessing GI endoscopes in simulated-use testing [J].
Foliente, RL ;
Kovacs, BJ ;
Aprecio, RM ;
Bains, HJ ;
Kettering, JD ;
Chen, YK .
GASTROINTESTINAL ENDOSCOPY, 2001, 53 (04) :456-462