A comprehensive physical image quality evaluation of a selenium based digital x-ray imaging system for thorax radiography

被引:32
作者
Launders, JH [1 ]
Kengyelics, SM [1 ]
Cowen, AR [1 ]
机构
[1] Gen Infirm, Dept Med Phys, FAXiL, Leeds LS1 3EX, W Yorkshire, England
关键词
digital radiography; amorphous selenium; image quality; chest radiography;
D O I
10.1118/1.598276
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A selenium based digital x-ray system dedicated to chest radiography has been installed by the UK Department of Health's Medical Devices Agency at Leeds General Infirmary, UK, to undergo a comprehensive evaluation, including the physical image quality. The underlying characteristics which define the overall image quality of a system are the following: sensitometric response, modulation transfer function, and noise power spectrum. These have been measured objectively on preprocessed digital data acquired under relevant radiographic conditions. The image data is further processed prior to hard copy display. The displayed image quality may only be measured subjectively; threshold contrast detail detectability is such a measure which can be related to the objective measures of image quality. The objective imaging characteristics suggest that Thoravision has a significant advantage over conventional radiography imaging systems. However, subjective measures have demonstrated that the image processing can have a significant effect on the perceived image quality. Thoravision has the potential to deliver a significantly improved image quality to clinicians with no increase in radiation exposure to the patient, or image quality may be maintained with a reduction in radiation exposure. Digital image processing is central to the efficiency with which it achieves this. (C) 1998 American Association of Physicists in Medicine. [S0094-2405(98)02706-0].
引用
收藏
页码:986 / 997
页数:12
相关论文
共 28 条
[11]   INVESTIGATION OF BASIC IMAGING PROPERTIES IN DIGITAL RADIOGRAPHY .2. NOISE WIENER SPECTRUM [J].
GIGER, ML ;
DOI, K ;
METZ, CE .
MEDICAL PHYSICS, 1984, 11 (06) :797-805
[12]   IMAGING PERFORMANCE OF A DIGITAL STORAGE PHOSPHOR SYSTEM [J].
HILLEN, W ;
SCHIEBEL, U ;
ZAENGEL, T .
MEDICAL PHYSICS, 1987, 14 (05) :744-751
[13]  
ICRU, 1996, 54 ICRU
[14]  
KENGYELICS SM, 1996, MDA9632 DEP HLTH
[15]   A COMPARISON OF THE THRESHOLD DETAIL DETECTABILITY OF A SCREEN-FILM COMBINATION AND COMPUTED RADIOLOGY UNDER CONDITIONS RELEVANT TO HIGH-KVP CHEST RADIOGRAPHY [J].
LAUNDERS, JH ;
COWEN, AR .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (08) :1393-1398
[16]  
LAUNDERS JH, 1996, MDA9631 DEP HLTH
[17]  
LAUNDERS JH, 1996, MDA9617 DEP HLTH
[18]  
McMaster W.H., 1969, Compilation of X-Ray Cross Sections, Lawerence Livermore Lab.
[19]   IMAGE QUALITY OF A DIGITAL CHEST RADIOGRAPHY SYSTEM BASED ON A SELENIUM DETECTOR [J].
NEITZEL, U ;
MAACK, I ;
GUNTHERKOHFAHL, S .
MEDICAL PHYSICS, 1994, 21 (04) :509-516
[20]  
NEITZEL U, 1995, COMMUNICATION