Transfer doublet and an elaborated phase plate holder for 120 kV electron-phase microscope

被引:32
作者
Hosokawa, F
Danev, R
Arai, Y
Nagayama, K
机构
[1] Okazaki Natl Res Inst, Ctr Integrat Biosci, Okazaki, Aichi 4448585, Japan
[2] JEOL Ltd, Akishima, Tokyo 1968558, Japan
来源
JOURNAL OF ELECTRON MICROSCOPY | 2005年 / 54卷 / 04期
关键词
transfer lens; phase plate; CTF; phase contrast;
D O I
10.1093/jmicro/dfi049
中图分类号
TH742 [显微镜];
学科分类号
摘要
A Zernike-type phase plate, which consisted of an amorphous carbon film, was developed for a 120 kV transmission electron microscope (TEM). The thickness of the film causes the scattered electrons to experience a pi/2 phase shift. The axial electrons pass through the central hole of the phase plate. The TEM was modified with an additional lens doublet, which allowed more flexibility in applying the phase plates. The doublet transfers the image from the back focal plane to a plane below the objective lens where a special phase plate holder, anticontaminator, and other necessary devices could be employed. The diameter of the central hole in the phase plate was 1 mm. The same mechanical system as in the specimen stage is used to align the axial beam into the hole. The use of a heating holder to protect the phase plate from contamination and consequential charging was investigated. Comparative diffractograms of amorphous carbon are shown with and without using a phase plate, demonstrating the change in phase contrast modulation from a sine to a cosine function when a phase plate is used. Images of biological specimens acquired using a phase plate are presented and the contrast enhancement due to the cosine transfer function is clearly demonstrated.A Zernike-type phase plate, which consisted of an amorphous carbon film, was developed for a 120 kV transmission electron microscope (TEM). The thickness of the film causes the scattered electrons to experience a pi/2 phase shift. The axial electrons pass through the central hole of the phase plate. The TEM was modified with an additional lens doublet, which allowed more flexibility in applying the phase plates. The doublet transfers the image from the back focal plane to a plane below the objective lens where a special phase plate holder, anticontaminator, and other necessary devices could be employed. The diameter of the central hole in the phase plate was 1 mu m. The same mechanical system as in the specimen stage is used to align the axial beam into the hole. The use of a heating holder to protect the phase plate from contamination and consequential charging was investigated. Comparative diffractograms of amorphous carbon are shown with and without using a phase plate, demonstrating the change in phase contrast modulation from a sine to a cosine function when a phase plate is used. Images of biological specimens acquired using a phase plate are presented and the contrast enhancement due to the cosine transfer function is clearly demonstrated.
引用
收藏
页码:317 / 324
页数:8
相关论文
共 12 条
[1]   A novel phase-contrast transmission electron microscopy producing high-contrast topographic images of weak objects [J].
Danev, R ;
Okawara, H ;
Usuda, N ;
Kametani, K ;
Nagayama, K .
JOURNAL OF BIOLOGICAL PHYSICS, 2002, 28 (04) :627-635
[2]   Complex observation in electron microscopy. II. Direct visualization of phases and amplitudes of exit wave functions [J].
Danev, R ;
Nagayama, K .
JOURNAL OF THE PHYSICAL SOCIETY OF JAPAN, 2001, 70 (03) :696-702
[3]   Transmission electron microscopy with Zernike phase plate [J].
Danev, R ;
Nagayama, K .
ULTRAMICROSCOPY, 2001, 88 (04) :243-252
[4]   Electric charging of thin films measured using the contrast transfer function [J].
Danov, K ;
Danev, R ;
Nagayama, K .
ULTRAMICROSCOPY, 2001, 87 (1-2) :45-54
[5]   RELATIVISTIC HARTREE-FOCK X-RAY AND ELECTRON SCATTERING FACTORS [J].
DOYLE, PA ;
TURNER, PS .
ACTA CRYSTALLOGRAPHICA SECTION A-CRYSTAL PHYSICS DIFFRACTION THEORETICAL AND GENERAL CRYSTALLOGRAPHY, 1968, A 24 :390-&
[6]  
FRANK J, 1973, OPTIK, V38, P519
[7]   NUMERICAL EVALUATION OF N-BEAM WAVE-FUNCTIONS IN ELECTRON-SCATTERING BY MULTI-SLICE METHOD [J].
GOODMAN, P ;
MOODIE, AF .
ACTA CRYSTALLOGRAPHICA SECTION A, 1974, A 30 (MAR) :280-290
[8]   Complex observation in electron microscopy. I. Basic scheme to surpass the Scherzer limit [J].
Nagayama, K .
JOURNAL OF THE PHYSICAL SOCIETY OF JAPAN, 1999, 68 (03) :811-822
[9]  
REIMER L, 1997, TRANSMISSION ELECT M, P51
[10]   THE THEORETICAL RESOLUTION LIMIT OF THE ELECTRON MICROSCOPE [J].
SCHERZER, O .
JOURNAL OF APPLIED PHYSICS, 1949, 20 (01) :20-29