Light output of Double Chooz scintillators for low energy electrons

被引:32
作者
Aberle, C. [1 ]
Buck, C. [1 ]
Hartmann, F. X. [1 ]
Schoenert, S. [1 ]
Wagner, S. [1 ]
机构
[1] Max Planck Inst Kernphys, D-69117 Heidelberg, Germany
来源
JOURNAL OF INSTRUMENTATION | 2011年 / 6卷
关键词
Scintillators; scintillation and light emission processes (solid; gas and liquid scintillators); Liquid detectors; Detector modelling and simulations I (interaction of radiation with matter; interaction of photons with matter; interaction of hadrons with matter; etc); Large detector systems for particle and astroparticle physics; ORGANIC MEDIA; RADIOLUMINESCENCE;
D O I
10.1088/1748-0221/6/11/P11006
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The energy dependent light output of liquid scintillators used in the Double Chooz experiment was measured for electrons up to 140 keV energy. A new Compton scattering coincidence apparatus was built for this purpose. A detailed study on possible systematic errors was made. We report the experimental results of our investigations and tested them for concordance with the predictions of various models. All models reasonably fit the experimental data after adjusting the respective free parameters. The results were also used to tune the Geant4-based Monte Carlo simulation software which is used in Double Chooz. The experimental data can be described by the simulation choosing an effective value for the Birks parameter.
引用
收藏
页数:19
相关论文
共 21 条
[1]   Light yield and energy transfer in a new Gd-loaded liquid scintillator [J].
Aberle, C. ;
Buck, C. ;
Hartmann, F. X. ;
Schoenert, S. .
CHEMICAL PHYSICS LETTERS, 2011, 516 (4-6) :257-262
[2]   GEANT4-a simulation toolkit [J].
Agostinelli, S ;
Allison, J ;
Amako, K ;
Apostolakis, J ;
Araujo, H ;
Arce, P ;
Asai, M ;
Axen, D ;
Banerjee, S ;
Barrand, G ;
Behner, F ;
Bellagamba, L ;
Boudreau, J ;
Broglia, L ;
Brunengo, A ;
Burkhardt, H ;
Chauvie, S ;
Chuma, J ;
Chytracek, R ;
Cooperman, G ;
Cosmo, G ;
Degtyarenko, P ;
Dell'Acqua, A ;
Depaola, G ;
Dietrich, D ;
Enami, R ;
Feliciello, A ;
Ferguson, C ;
Fesefeldt, H ;
Folger, G ;
Foppiano, F ;
Forti, A ;
Garelli, S ;
Giani, S ;
Giannitrapani, R ;
Gibin, D ;
Cadenas, JJG ;
González, I ;
Abril, GG ;
Greeniaus, G ;
Greiner, W ;
Grichine, V ;
Grossheim, A ;
Guatelli, S ;
Gumplinger, P ;
Hamatsu, R ;
Hashimoto, K ;
Hasui, H ;
Heikkinen, A ;
Howard, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 506 (03) :250-303
[3]  
[Anonymous], 1964, THEORY PRACTICE SCIN
[4]  
Ardellier F, hep-ex/0606025.
[5]   SCINTILLATIONS FROM ORGANIC CRYSTALS - SPECIFIC FLUORESCENCE AND RELATIVE RESPONSE TO DIFFERENT RADIATIONS [J].
BIRKS, JB .
PROCEEDINGS OF THE PHYSICAL SOCIETY OF LONDON SECTION A, 1951, 64 (382) :874-877
[6]   Energy transfer and light yield properties of a new highly loaded indium(III) β-diketonate organic scintillator system [J].
Buck, C. ;
Hartmann, F. X. ;
Motta, D. ;
Schoenert, S. .
CHEMICAL PHYSICS LETTERS, 2007, 435 (4-6) :252-256
[7]   Luminescent properties of a new In-based organic liquid scintillation system [J].
Buck, C ;
Hartmann, F ;
Lasserre, T ;
Motta, D ;
Schönert, S ;
Schwan, U .
JOURNAL OF LUMINESCENCE, 2004, 106 (01) :57-67
[8]   Development of an optically pure In β-diketonate for the scintillator of an 115In-loaded solar neutrino detector [J].
Buck, C ;
Hartmann, FX ;
Schönert, S ;
Schwan, U .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2003, 258 (02) :255-263
[9]  
BUCK C, J RADIOANAL NU UNPUB
[10]   THE NATURE OF THE SATURATION EFFECT OF FLUORESCENT SCINTILLATORS [J].
CHOU, CN .
PHYSICAL REVIEW, 1952, 87 (05) :904-905