Emboli capture using the Embol-X IntraAortic filter in cardiac surgery: A multicentered randomized trial of 1,289 patients

被引:60
作者
Banbury, MK
Kouchoukos, NT
Allen, KB
Slaughter, MS
Weissman, NJ
Berry, GJ
Horvath, KA
机构
[1] Cleveland Clin Fdn, Cleveland, OH 44195 USA
[2] Missouri Baptist Med Ctr, St Louis, MO USA
[3] St Vincents Med Ctr, Indianapolis, IN USA
[4] Christ Hosp, Oak Lawn, IL USA
[5] Washington Hosp Ctr, Washington, DC 20010 USA
[6] Stanford Univ, Ctr Med, Stanford, CA USA
[7] Northwestern Univ, Ctr Med, Chicago, IL USA
关键词
D O I
10.1016/S0003-4975(03)00530-7
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Particulate emboli are thought to play a significant role in the development of cardiac surgical complications. Intraaortic filtration of particulate emboli may reduce the burden of this morbidity in cardiac patients. Methods. A multiinstitutional randomized trial was designed and enrolled 1,289 patients at 22 centers. Six hundred forty-five patients were assigned to the treatment arm and received the Embol-X intraaortic filter, whereas 644 patients were assigned to the control arm. The endpoints examined were mortality, stroke, transient ischemic attack, renal insufficiency/failure, myocardial infarction, gastrointestinal complications, and limb-threatening ischemia. All filters were examined for histologic evidence of particulate emboli. Results. Particulate emboli were identified in 598 (96.8%) of 618 filters successfully deployed. Composite event rates for the clinical endpoints were similar in both the filtered and nonfiltered arm (110/645 = 17% vs 122/644 = 19%, respectively). Individual event rates were also similar in both arms. A post hoc comparison of patients with moderate or greater preoperative risk scores demonstrated event reduction favoring the filtered group for renal complications (17/124 = 14% vs 28/117 = 24%, p = 0.04) and for the composite endpoint (30/124 = 24% vs 42/117 = 36%, p = 0.047). No clinically evident complications attributed to the use of the filter were identified. Conclusions. The use of the Embol-X intraaortic filter is both safe and effective, as demonstrated by the emboli capture rate of 97%. In addition, post hoc analysis indicates a reduction in postoperative renal complications for patients with moderate or greater preoperative risk. Further study of high-risk patients is warranted. (C) 2003 by The Society of Thoracic Surgeons.
引用
收藏
页码:508 / 515
页数:8
相关论文
共 20 条
[11]   IMPROVED RELIABILITY OF THE NIH STROKE SCALE USING VIDEO TRAINING [J].
LYDEN, P ;
BROTT, T ;
TILLEY, B ;
WELCH, KMA ;
MASCHA, EJ ;
LEVINE, S ;
HALEY, EC ;
GROTTA, J ;
MARLER, J .
STROKE, 1994, 25 (11) :2220-2226
[12]   MULTIPLE TESTING PROCEDURE FOR CLINICAL-TRIALS [J].
OBRIEN, PC ;
FLEMING, TR .
BIOMETRICS, 1979, 35 (03) :549-556
[13]   Particulate emboli capture by an intra-aortic filter device during cardiac surgery [J].
Reichenspurner, H ;
Navia, JA ;
Berry, G ;
Robbins, RC ;
Barbut, D ;
Gold, JP ;
Reichart, B .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2000, 119 (02) :233-240
[14]   Predictors of outcome in cardiac surgical patients with prolonged intensive care stay [J].
Ryan, TA ;
Rady, MY ;
Bashour, CA ;
Leventhal, M ;
Lytle, B ;
Starr, NJ .
CHEST, 1997, 112 (04) :1035-1042
[15]   International Council of Emboli Management (ICEM) Study Group results: risk adjusted outcomes in intraaortic filtration [J].
Schmitz, C ;
Blackstone, EH .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2001, 20 (05) :986-990
[16]  
STUMP DA, 1996, NEUROSONOLOGY, P252
[17]   Recognition of the importance of embolization in atherosclerotic vascular disease [J].
Topol, EJ ;
Yadav, JS .
CIRCULATION, 2000, 101 (05) :570-580
[18]   Extracorporeal circulation before and after ultrasonographic evaluation of the ascending aorta [J].
Ura, M ;
Sakata, R ;
Nakayama, Y ;
Miyamoto, TA ;
Goto, T .
ANNALS OF THORACIC SURGERY, 1999, 67 (02) :478-483
[19]   Cognitive outcome after off-pump and on-pump coronary artery bypass graft surgery - A randomized trial [J].
Van Dijk, D ;
Jansen, EWL ;
Hijman, R ;
Nierich, AP ;
Diephuis, JC ;
Moons, KGM ;
Lahpor, JR ;
Borst, C ;
Keizer, AMA ;
Nathoe, HM ;
Grobbee, DE ;
De Jaegere, PPT ;
Kalkman, CJ .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2002, 287 (11) :1405-1412
[20]  
WAREING TH, 1992, J THORAC CARDIOV SUR, V103, P453