Dysfunction of implantable defibrillators caused by slot machines

被引:26
作者
Madrid, A
Sanchez, A
Bosch, E
Fernandez, E
Serrano, CM
机构
[1] UNIV ALCALA DE HENARES, HOSP RAMON Y CAJAL, ARRHYTHMIS UNIT, DEPT MED, MADRID, SPAIN
[2] UNIV ALCALA DE HENARES, HOSP RAMON Y CAJAL, CPI GUIDANT ENG, DEPT MED, MADRID, SPAIN
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1997年 / 20卷 / 01期
关键词
defibrillators; electromagnetic interference; slot machines;
D O I
10.1111/j.1540-8159.1997.tb04845.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The electrogram storage is a feature offered by the last generation of defibrillators, which enables the physician to establish whether the sensed episode and the electrical therapy delivered by the device was appropriate. The incidence of electromagnetic interferences may be higher than initially suspected with the latest generation of defibrillators. Such interference was documented in 4 out of 100 patients over a 24-month period. The electrical interference occurred in all four of them, while they were playing with slot machines. In the RR interval history, irregular QRS sensing of variable frequency was documented in some episodes. The intervals were irregular and nonphysiological, satisfying the rate cut-off criteria in an intermittent and nonsustained way. There may be high risk of electromagnetic interference between slot machines and defibrillators. Our experience calls for a warning to all defibrillator patients and, perhaps some regulatory intervention.
引用
收藏
页码:212 / 214
页数:3
相关论文
共 10 条
[1]   ADVANCES IN FOLLOW-UP TECHNIQUES FOR IMPLANTABLE DEFIBRILLATORS [J].
BRACHMANN, J ;
STERNS, LD ;
HILBEL, T ;
BEYER, T ;
SCHOELS, W ;
FREIGANG, K ;
MELICHERCIK, J ;
RUFRICHTER, J ;
KUBLER, W .
AMERICAN HEART JOURNAL, 1994, 127 (04) :1081-1085
[2]   THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - EVALUATING SUSPECTED INAPPROPRIATE SHOCKS [J].
CHAPMAN, PD ;
TROUP, P .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1986, 7 (05) :1075-1078
[3]   ELECTRONIC ARTICLE SURVEILLANCE - A POSSIBLE DANGER FOR PACEMAKER PATIENTS [J].
DODINOT, B ;
GODENIR, JP ;
COSTA, AB ;
ZELLER, C ;
BROSCHART, M .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (01) :46-53
[4]   POTENTIAL HAZARDS OF IMPLANTED DEVICES FOR THE ELECTRICAL CONTROL OF TACHYARRHYTHMIAS [J].
ECHT, DS .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1984, 7 (03) :580-587
[5]   IMPORTANCE OF ABORTIVE SHOCK CAPABILITY WITH ELECTROGRAM STORAGE IN CARDIOVERTER-DEFIBRILLATOR DEVICES [J].
HURWITZ, JL ;
HOOK, BG ;
FLORES, BT ;
MARCHLINSKI, FE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 21 (04) :895-900
[6]   ARRHYTHMIAS INDUCED BY DEVICE ANTITACHYCARDIA THERAPY DUE TO DIAGNOSTIC NONSPECIFICITY [J].
JOHNSON, NJ ;
MARCHLINSKI, FE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 18 (05) :1418-1425
[7]   UNIPOLAR PACER ARTIFACTS INDUCED FAILURE OF AN AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR TO DETECT VENTRICULAR-FIBRILLATION [J].
KIM, SG ;
FURMAN, S ;
WASPE, LE ;
BRODMAN, R ;
FISHER, JD .
AMERICAN JOURNAL OF CARDIOLOGY, 1986, 57 (10) :880-881
[8]   ICD DATA-STORAGE - VALUE IN ARRHYTHMIA MANAGEMENT [J].
MARCHLINSKI, FE ;
GOTTLIEB, CD ;
SARTER, B ;
FINKLE, J ;
HOOK, B ;
CALLANS, D ;
SCHWARTZMAN, D .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (03) :527-534
[9]   THE AUTOMATIC IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - EFFICACY, COMPLICATIONS, AND DEVICE FAILURES [J].
MARCHLINSKI, FE ;
FLORES, BT ;
BUXTON, AE ;
HARGROVE, WC ;
ADDONIZIO, VP ;
STEPHENSON, LW ;
HARKEN, AH ;
DOHERTY, JU ;
GROGAN, EW ;
JOSEPHSON, ME .
ANNALS OF INTERNAL MEDICINE, 1986, 104 (04) :481-488
[10]   INAPPROPRIATE DISCHARGE BY AN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR - RECOGNITION OF MYOPOTENTIAL SENSING USING TELEMETERED INTRACARDIAC ELECTROGRAMS [J].
SANDLER, MT ;
KUTALEK, SP .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (04) :665-671