Development of a pivot bearing supported sealless centrifugal pump for ventricular assist

被引:29
作者
Nakazawa, T
Makinouchi, K
Ohara, Y
Ohtsubo, S
Kawahito, K
Tasai, K
Shimono, T
Benkowski, R
Damm, G
Takami, Y
Glueck, J
Noon, GP
Nose, Y
机构
[1] Department of Surgery, Baylor College of Medicine, Houston, TX
[2] Department of Surgery, Baylor College of Medicine, Houston, TX 77030, One Baylor Plaza
关键词
pivot bearing; centrifugal pump; ventricular assist device; totally implantable pump;
D O I
10.1111/j.1525-1594.1996.tb04469.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Since 1991, in our laboratory, a pivot bearing-supported, sealless, centrifugal pump has been developed as an implantable ventricular assist device (VAD). For this application, the configuration of the total pump system should be relatively small. The C1E3 pump developed for this purpose was anatomically compatible with the small-sized patient population. To evaluate antithrombogenicity, ex vivo 2-week screening studies were conducted instead of studies involving an intracorporeally implanted VADs using calves. Five paracorporeal LVAD studies were performed using calves for longer than 2 weeks. The activated clotting time (ACT) was maintained at approximately 250 s using heparin. All of the devices demonstrated trouble-free performances over 2 weeks. Among these 5 studies, 3 implantations were subjected to 1-month system validation studies. There were no device-induced thrombus formations inside the pump housing, and plasma-free hemoglobin levels in calves were within the normal range throughout the experiment (35, 34, and 31 days). There were no incidents of system malfunction. Subsequently, the mass production model was fabricated and yielded a normalized index of hemolysis of 0.0014, which was comparable to that of clinically available pumps. The wear life of the impeller bearings was estimated at longer than 8 years. In the next series of in vivo studies, an implantable model of the C1E3 pump will be fabricated for longer term implantation. The pump-actuator will be implanted inside the body; thus the design calls for substituting plastic for metallic parts.
引用
收藏
页码:485 / 490
页数:6
相关论文
共 15 条
[1]   A FLOW VISUALIZATION STUDY OF THE NCVC CENTRIFUGAL BLOOD PUMP [J].
ARAKI, K ;
TAENAKA, Y ;
MASUZAWA, T ;
TATSUMI, E ;
WAKISAKA, Y ;
WATARI, M ;
NAKATANI, T ;
AKAGI, H ;
BABA, Y ;
ANAI, H ;
PARK, YH ;
EYA, K ;
TODA, K ;
TAKANO, H .
ARTIFICIAL ORGANS, 1994, 18 (09) :669-672
[2]   FREQUENCY OF SEAL DISRUPTION WITH THE SARNS CENTRIFUGAL PUMP IN POSTCARDIOTOMY CIRCULATORY ASSIST [J].
CURTIS, JJ ;
BOLEY, TM ;
WALLS, JT ;
DEMMY, TL ;
SCHMALTZ, RA .
ARTIFICIAL ORGANS, 1994, 18 (03) :235-237
[3]   INITIAL CLINICAL-EXPERIENCE WITH A NEW TEMPORARY LEFT-VENTRICULAR ASSIST DEVICE [J].
GOLDING, LR ;
GROVES, LK ;
PETER, M ;
JACOBS, G ;
SUKALAC, R ;
NOSE, Y ;
LOOP, FD .
ANNALS OF THORACIC SURGERY, 1980, 29 (01) :66-69
[4]   SYSTEM CONSIDERATIONS FAVORING ROTARY ARTIFICIAL HEARTS WITH BLOOD-IMMERSED BEARINGS [J].
JARVIK, RK .
ARTIFICIAL ORGANS, 1995, 19 (07) :565-570
[5]   Evaluation of the wear of the pivot bearing in the Gyro C1E3 pump [J].
Makinouchi, K ;
Nakazawa, T ;
Takami, Y ;
Takatani, S ;
Nose, Y .
ARTIFICIAL ORGANS, 1996, 20 (06) :523-528
[6]  
MINATO N, 1993, ARTIF ORGANS, V17, P36
[7]   Modification of a pivot bearing system on a compact centrifugal pump [J].
Nakazawa, T ;
Makinouchi, K ;
Takami, Y ;
Glueck, J ;
Takatani, S ;
Nose, Y .
ARTIFICIAL ORGANS, 1996, 20 (03) :258-263
[8]   The effect of the impeller-driver magnetic coupling distance on hemolysis in a compact centrifugal pump [J].
Nakazawa, T ;
Makinouchi, K ;
Takami, Y ;
Glueck, J ;
Takatani, S ;
Nose, Y .
ARTIFICIAL ORGANS, 1996, 20 (03) :252-257
[9]   CAN WE DEVELOP A TOTALLY IMPLANTABLE ROTARY BLOOD PUMP [J].
NOSE, Y .
ARTIFICIAL ORGANS, 1995, 19 (07) :561-562
[10]   DEVELOPMENT AND EVALUATION OF ANTITHROMBOGENIC CENTRIFUGAL PUMP - THE BAYLOR-C-GYRO-PUMP-ECCENTRIC-INLET-PORT-MODEL [J].
OHARA, Y ;
MAKINOUCHI, K ;
GLUECK, J ;
SUTHERLAND, B ;
SHIMONO, T ;
NAITO, K ;
TASAI, K ;
ORIME, Y ;
TAKATANI, S ;
NOSE, Y .
ARTIFICIAL ORGANS, 1994, 18 (09) :673-679