Explosives Detection Using Direct Analysis in Real Time (DART) Mass Spectrometry

被引:102
作者
Nilles, J. Michael [1 ]
Connell, Theresa R. [1 ]
Stokes, Sarah T. [1 ]
Durst, H. Dupont [2 ]
机构
[1] Excet Inc, Springfield, VA 22151 USA
[2] Aberdeen Proving Ground, Edgewood Chem Biol Ctr, Edgewood, MD 21040 USA
关键词
DESORPTION ELECTROSPRAY-IONIZATION; ATMOSPHERIC PRESSURE PHOTOIONIZATION; ION MOBILITY SPECTROMETRY; CHEMICAL WARFARE AGENTS; OF-FLIGHT; TRACE ANALYSIS; RAPID ANALYSIS; IDENTIFICATION; SAMPLES; ACID;
D O I
10.1002/prep.200900084
中图分类号
O69 [应用化学];
学科分类号
081704 ;
摘要
The growing use of explosives by terrorists and criminals creates a need for instrumentation which can rapidly analyze these energetic compounds, preferably on site. Direct analysis in real time (DART) is a promising technology for surface analysis with little or no sample preparation. Therefore, DART ionization is evaluated for use in detecting explosives on solid substrates and in liquid matrices. Fifteen explosives were chosen as a consequence of their common usage. Five surfaces were chosen to represent a wide range of physical properties such as composition, porosity, surface morphology, and thermal and electrical conductivity. Additionally these surfaces are commonly found in everyday surroundings. All 75 compound-surface combinations produced a clear, easily identifiable, mass spectra characteristic of the targeted analyte. Simultaneous detection of five explosives is demonstrated on these same surfaces. Lastly, rapid detection of trace contamination in common fluids is also explored.
引用
收藏
页码:446 / 451
页数:6
相关论文
共 43 条
[31]   Trace analysis of peroxide-based explosives [J].
Schulte-Ladbeck, R ;
Kolla, P ;
Karst, U .
ANALYTICAL CHEMISTRY, 2003, 75 (04) :731-735
[32]   Control of Strobilurin Fungicides in Wheat Using Direct Analysis in Real Time Accurate Time-of-Flight and Desorption Electrospray Ionization Linear Ion Trap Mass Spectrometry [J].
Schurek, Jakub ;
Vaclavik, Lukas ;
Hooijerink, H. ;
Lacina, Ondrej ;
Poustka, Jan ;
Sharman, Matthew ;
Caldow, Marianne ;
Nielen, Michel W. F. ;
Hajslova, Jana .
ANALYTICAL CHEMISTRY, 2008, 80 (24) :9567-9575
[33]   Analysis of triacetone triperoxide by gas chromatography/mass spectrometry and gas chromatography/tandem mass spectrometry by electron and chemical ionization [J].
Sigman, Michael E. ;
Clark, C. Douglas ;
Fidler, Rebecca ;
Geiger, Cherie L. ;
Clausen, Christian A. .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2006, 20 (19) :2851-2857
[34]   HRMS directly from TLC slides. A powerful tool for rapid analysis of organic mixtures [J].
Smith, Natalie J. ;
Domin, Marek A. ;
Scott, Lawrence T. .
ORGANIC LETTERS, 2008, 10 (16) :3493-3496
[35]   Negative ion-atmospheric pressure photoionization: Electron capture, dissociative electron capture, proton transfer, and anion attachment [J].
Song, Liguo ;
Wellman, Amber D. ;
Yao, Huifang ;
Bartmess, John E. .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2007, 18 (10) :1789-1798
[36]   Ionization Mechanism of Negative Ion-Direct Analysis in Real Time: A Comparative Study with Negative Ion-Atmospheric Pressure Photoionization [J].
Song, Liguo ;
Dykstra, Andrew B. ;
Yao, Huifang ;
Bartmess, John E. .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2009, 20 (01) :42-50
[37]   Validation of the Direct Analysis in Real Time Source for Use in Forensic Drug Screening [J].
Steiner, Robert R. ;
Larson, Robyn L. .
JOURNAL OF FORENSIC SCIENCES, 2009, 54 (03) :617-622
[38]  
VACLAVIK L, 2008, CHEM LISTY, V102, pS324
[39]   Rapid and unambiguous identification of melamine in contaminated pet food based on mass spectrometry with four degrees of confirmation [J].
Vail, Teresa M. ;
Jones, Patrick R. ;
Sparkman, O. David .
JOURNAL OF ANALYTICAL TOXICOLOGY, 2007, 31 (06) :304-312
[40]   Development of an LC/MS method for the trace analysis of triacetone triperoxide (TATP) [J].
Widmer, L ;
Watson, S ;
Schlatter, K ;
Crowson, A .
ANALYST, 2002, 127 (12) :1627-1632