"Thicker" Polyethylene Bearings Are Associated With Higher Failure Rates in Primary Total Knee Arthroplasty

被引:51
作者
Berend, Michael E. [1 ]
Davis, Peter J. [1 ]
Ritter, Merrill A. [1 ]
Keating, E. Michael [1 ]
Faris, Philip M. [1 ]
Meding, John B. [1 ]
Malinzak, Robert A. [1 ]
机构
[1] St Francis Hosp Mooresville, Joint Replacement Surg Indiana Res Fdn, Ctr Hip & Knee Surg, St Francis Med Grp, Mooresville, IN 46158 USA
关键词
total knee arthroplasty; polyethylene bearing; thickness; failure rate; IMPLANT MIGRATION; REPLACEMENT; COMPONENTS; WEAR; THICKNESS; BONE;
D O I
10.1016/j.arth.2010.04.031
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
"Thin" modular polyethylene bearings have previously been associated with failure from wear. This study examined the influence of polyethylene thickness on survivorship in primary total knee arthroplasty (TKA). Do "thinner" or "thicker" bearings fail more? Six thousand seventy primary TKAs with a single implant design were reviewed. The failure rate in knees with bearings 14 mm or less was 0.7%, whereas the failure rate of knees with bearings 16 mm or greater was 2.3% (P < .0001; hazard ratio, 3.2). No knee was revised for polyethylene wear. Thicker bearings did not directly cause failure, but factors that lead to the insertion of a thicker bearing such as a deeper tibial resection and ligament imbalance may contribute to the observed increased failure. The significant influence of this often-unrecognized surgical variable has not been previously reported and must be carefully considered during TKA.
引用
收藏
页码:17 / 20
页数:4
相关论文
共 18 条
[1]
THE EFFECT OF CONFORMITY, THICKNESS, AND MATERIAL ON STRESSES IN ULTRAHIGH MOLECULAR-WEIGHT COMPONENTS FOR TOTAL JOINT REPLACEMENT [J].
BARTEL, DL ;
BICKNELL, VL ;
WRIGHT, TM .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1986, 68A (07) :1041-1051
[2]
Tibial component failure mechanisms in total knee arthroplasty [J].
Berend, ME ;
Ritter, MA ;
Meding, JB ;
Faris, PM ;
Keating, EM ;
Redelman, R ;
Faris, GW ;
Davis, KE .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (428) :26-34
[3]
Implant migration and failure in total knee arthroplasty is related to body mass index and tibial component size [J].
Berend, Michael E. ;
Ritter, Merrill A. ;
Hyldahl, Hans Christian ;
Meding, John B. ;
Redelman, Ryan .
JOURNAL OF ARTHROPLASTY, 2008, 23 (06) :104-109
[4]
The Effects of Bone Resection Depth and Malalignment on Strain in the Proximal Tibia After Total Knee Arthroplasty [J].
Berend, Michael E. ;
Small, Scott R. ;
Ritter, Merrill A. ;
Buckley, Christine A. .
JOURNAL OF ARTHROPLASTY, 2010, 25 (02) :314-318
[5]
Chaput Christopher D, 2004, J Surg Orthop Adv, V13, P195
[6]
Effect of ultra-high molecular weight polyethylene thickness on contact mechanics in total knee replacement [J].
El-Deen, M. ;
Garcia-Finana, M. ;
Jin, Z-M .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2006, 220 (H7) :733-742
[7]
POLYETHYLENE WEAR OF METAL-BACKED TIBIAL COMPONENTS IN TOTAL AND UNICOMPARTMENTAL KNEE PROSTHESES [J].
ENGH, GA ;
DWYER, KA ;
HANES, CK .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (01) :9-17
[8]
EWALD FC, 1989, CLIN ORTHOP RELAT R, P9
[9]
The effects of varus tibial alignment on proximal tibial surface strain in total knee arthroplasty - The posteromedial hot spot [J].
Green, GV ;
Berend, KR ;
Berend, ME ;
Glisson, RR ;
Vail, TP .
JOURNAL OF ARTHROPLASTY, 2002, 17 (08) :1033-1039
[10]
HOFMANN AA, 1991, CLIN ORTHOP RELAT R, P63