Confirmatory and quantitative analysis of β-lactam antibiotics in bovine kidney tissue by dispersive solid-phase extraction and liquid chromatography-tandem mass spectrometry

被引:122
作者
Fagerquist, CK [1 ]
Lightfield, AR
Lehotay, SJ
机构
[1] USDA ARS, Western Reg Res Ctr, Albany, CA 94710 USA
[2] USDA ARS, Eastern Reg Res Ctr, Wyndmoor, PA 19038 USA
关键词
D O I
10.1021/ac040138q
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple, rapid, rugged, sensitive, and specific method for the confirmation and quantitation of 10 beta-lactam antibiotics in fortified and incurred bovine kidney tissue has been developed. The method uses a simple solvent extraction, dispersive solid-phase extraction (dispersive SPE) cleanup, and liquid chromatography-tandem mass spectrometry (LC/MS/MS) for confirmation and quantitation. Dispersive-SPE greatly simplifies and accelerates sample cleanup and improves overall recoveries compared with conventional SPE cleanup. The beta-lactam antibiotics tested were as follows: deacetylcephapirin (an antimicrobial metabolite of cephapirin), amoxicillin, desfuroylceftiofur cysteine disulfide (DCCD, an antimicrobial metabolite of ceftiofur), ampicillin, cefazolin, penicillin G, oxacillin, cloxacillin, naficillin, and dicloxacillin. Average recoveries of fortified samples were 70% or better for all beta-lactams except DCCD, which had an average recovery of 58%. The LC/MS/MS method was able to demonstrate quantitative recoveries at established tolerance levels and provide confirmatory data for unambiguous analyte identification. The method was also tested on 30 incurred bovine kidney samples obtained from the USDA Food Safety and Inspection Service, which had previously tested the samples using the approved semiquantitative microbial assay. The results from the quantitative LC/MS/MS analysis were in general agreement with the microbial assay for 23 samples although the LC/MS/MS method was superior in that it could specifically identify which beta-lactam was present and quantitate its concentration, whereas the microbial assay could only identify the type of beta-lactam present and report a concentration with respect to the microbial inhibition of a penicillin G standard. In addition, for 6 of the 23 samples, LC/MS/ MS analysis detected a penicillin and a cephalosporin beta-lactam, whereas the microbial assay detected only a penicillin beta-lactam. For samples that do not fall into the "general agreement" category, the most serious discrepancy involves two samples where the LC/MS/MS method detected a violative level of a cephalosporin beta-lactam (deacetylcephapirin) in the first sample and a possibly violative level of desfuroylceftiofur in the second, whereas the microbial assay identified the two samples as having only violative levels of a penicillin beta-lactam.
引用
收藏
页码:1473 / 1482
页数:10
相关论文
共 33 条
[11]   Confirmatory analysis of β-lactam antibiotics in kidney tissue by liquid chromatography/electrospray ionization selective reaction monitoring ion trap tandem mass spectrometry [J].
Fagerquist, CK ;
Lightfield, AR .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2003, 17 (07) :660-671
[12]   An antibiotic linked to peptides and proteins is released by electron capture dissociation Fourier transform ion cyclotron resonance mass spectrometry [J].
Fagerquist, CK ;
Hudgins, RR ;
Emmett, MR ;
Håkansson, K ;
Marshall, AG .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2003, 14 (04) :302-310
[13]  
Heller DN, 1998, RAPID COMMUN MASS SP, V12, P2031, DOI 10.1002/(SICI)1097-0231(19981230)12:24<2031::AID-RCM432>3.0.CO
[14]  
2-8
[15]   Screening and mass spectral confirmation of β-lactam antibiotic residues in milk using LC-MS/MS [J].
Holstege, DM ;
Puschner, B ;
Whitehead, G ;
Galey, FD .
JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 2002, 50 (02) :406-411
[16]  
Hornish RE, 2003, J AOAC INT, V86, P30
[17]   Application of ion-exchange cartridge clean-up in food analysis IV. Confirmatory assay of benzylpenicillin, phenoxymethylpenicillin, oxacillin, cloxacillin, nafcillin and dicloxacillin in bovine tissues by liquid chromatography-electrospray ionization tandem mass spectrometry [J].
Ito, Y ;
Ikai, Y ;
Oka, H ;
Matsumoto, H ;
Miyazaki, Y ;
Takeba, K ;
Nagase, H .
JOURNAL OF CHROMATOGRAPHY A, 2001, 911 (02) :217-223
[18]  
JAGLAN PS, 1990, J ASSOC OFF ANA CHEM, V73, P26
[19]   Quantitative determination of ceftiofur in milk by liquid chromatography electrospray mass spectrometry [J].
Keever, J ;
Voyksner, RD ;
Tyczkowska, KL .
JOURNAL OF CHROMATOGRAPHY A, 1998, 794 (1-2) :57-62
[20]   Acceptance criteria for ultratrace HPLC-tandem mass spectrometry: Quantitative and qualitative determination of sulfonylurea herbicides in soil [J].
Li, LYT ;
Campbell, DA ;
Bennett, PK ;
Henion, J .
ANALYTICAL CHEMISTRY, 1996, 68 (19) :3397-3404