The effect of hydroxyapatite on the micromotion of total knee prostheses - A prospective, randomized, double-blind study

被引:94
作者
Nelissen, RGHH [1 ]
Valstar, ER [1 ]
Rozing, PM [1 ]
机构
[1] Leiden Univ, Med Ctr, Dept Orthopaed Surg, NL-2300 RC Leiden, Netherlands
关键词
D O I
10.2106/00004623-199811000-00014
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
A prospective, randomized, double-blind study was performed to evaluate three different means of fixing tibial components during total knee arthroplasty, Eleven components fixed with cement, ten hydroxyapatite-coated components fixed without cement, and ten noncoated components fixed without cement were studied, A posterior cruciate ligament-retaining total condylar implant was used. Micromotion of the components was assessed with roentgen stereophotogrammetric analysis during the two-year follow-up period. There were no significant differences among the patients with regard to age (mean [and standard deviation], 68 +/- 11.6 years), body-mass index (mean, 23 +/- 2.8 kilograms per square meter), or stage of osteoarthrosis (mean, 4 +/- 2.4 according to the classification system of Ahlback and 5 +/- 0.6 according to that of Larsen et al,), The diagnosis was osteoarthrosis in five knees, and it was rheumatoid arthritis in twenty-six, The clinical scores were similar among the study groups. According to the system of the Knee Society, the mean preoperative functional score was 10 +/- 2.9 points and the mean preoperative knee score was 24 +/- 3.2 points. At the two-year follow-up evaluation, these scores were 41 +/- 8.3 and 79 +/- 3.2 points, respectively. A significant difference with regard to micromotion was found between the noncoated components fixed without cement and the hydroxyapatite-coated components fixed without cement as well as between the noncoated components fixed without cement and the components fixed with cement (p < 0.001, analysis of variance). The hydroxyapatite-coated components fixed without cement and the components fixed with cement both had far less micromotion along the longitudinal axis (subsidence) throughout the follow-up period than did the noncoated components fixed without cement. At the two-year follow-up evaluation, the subsidence of the noncoated components was -0.73 +/- 0.924 millimeter, the subsidence of the cemented components was -0.05 +/- 0.109 millimeter, and the subsidence of the hydroxyapatite-coated components was -0.06 +/- 0.169 millimeter. The cemented components as well as the hydroxyapatite-coated components also had less translation along the transverse axis (p < 0.001, analysis of variance) and the sagittal axis (p < 0.001, analysis of variance) compared with the noncoated components. In conclusion, micromotion of hydroxyapatite-coated tibial components fixed without cement was similar to that of tibial components fixed with cement. Therefore, hydroxyapatite, a biological mediator, may be necessary for the adequate fixation of tibial components when cement is not used.
引用
收藏
页码:1665 / 1672
页数:8
相关论文
共 27 条
[1]
AHLBACK S, 1968, ACTA RADIOL DIAG S, V277
[2]
BLOEBAUM RD, 1994, CLIN ORTHOP RELAT R, P19
[3]
BUECHEL FF, 1994, ORTHOPEDICS, V17, P833
[4]
BUECHEL FF, 1989, ORTHOP CLIN N AM, V20, P147
[5]
EWALD FC, 1989, CLIN ORTHOP RELAT R, P9
[6]
FREEMAN MAR, 1992, CLIN ORTHOP RELAT R, V276, P19
[7]
6-YEAR RESULTS OF HYDROXYAPATITE-COATED TOTAL HIP-REPLACEMENT [J].
GEESINK, RGT ;
HOEFNAGELS, NHM .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1995, 77B (04) :534-547
[8]
THE STABILITY OF 3 DIFFERENT CEMENTLESS TIBIAL COMPONENTS - A RANDOMIZED RADIOSTEREOMETRIC STUDY IN 45 KNEE ARTHROPLASTY PATIENTS [J].
HILDING, MB ;
YUAN, XH ;
RYD, L .
ACTA ORTHOPAEDICA SCANDINAVICA, 1995, 66 (01) :21-27
[9]
Insall J. N., 1993, SURG KNEE
[10]
DOES EARLY MICROMOTION OF FEMORAL STEM PROSTHESES MATTER - 4-7-YEAR STEREORADIOGRAPHIC FOLLOW-UP OF 84 CEMENTED PROSTHESES [J].
KARRHOLM, J ;
BORSSEN, B ;
LOWENHIELM, G ;
SNORRASON, F .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (06) :912-917