Dislodgment of polyethylene liners in first and second-generation Harris-Galante acetabular components - A report of eighteen cases

被引:69
作者
Della Valle, AG
Ruzo, PS
Li, S
Pellicci, P
Sculco, TP
Salvati, EA
机构
[1] Hosp Special Surg, Dept Biomech, New York, NY 10021 USA
[2] Hosp Special Surg, Dept Orthoped Surg, New York, NY 10021 USA
关键词
D O I
10.2106/00004623-200104000-00010
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Dislodgment of the polyethylene liner is an increasingly common complication following total hip arthroplasty. The purposes of this study are to present the results in a series of patients with this complication and to analyze the mode of failure. Methods: Between November 1995 and January 2001, eighteen patients who had had a total hip arthroplasty presented with dislodgment of the polyethylene liner from a Harris-Galante metal acetabular shell. The medical records, radiographs, operative notes, and retrieved components were reviewed. In addition, scanning electron microscopy was used to study the fractured surfaces in a shell that had four broken tines. Results: The components had been in situ for an average of seven years (range, three to eleven years). Seventeen components were second generation, and one was first generation. Symptoms developed spontaneously in sixteen patients, during sexual intercourse in one, and following a fall on the hip in one. Radiographs showed eccentric positioning of the head in all of the hips and broken tines in six. All of the shells were well fixed. Treatment consisted of revision of the shell in four patients, exchange of the liner in four, cementation of a new liner into the shell in seven, and cementation of an all-polyethylene cup in three. The liners had severe damage of the rim. Scanning-electron microscopy of the fractured surfaces of four tines revealed a fatigue pattern. Conclusions: We believe that, as the liner wears and becomes loose because of an inadequate locking mechanism, progressive micromotion occurs and the load increases on the polyethylene rim until it deforms and/or fractures. Subsequently, nothing prevents the liner from rotating out of the shell. As this mechanism of failure appears to include fatigue failure of the locking tines and wear of the liner, this complication is likely to increase as the components age in situ.
引用
收藏
页码:553 / 559
页数:7
相关论文
共 44 条
[1]
ANTEVERSION OF THE ACETABULAR CUP - MEASUREMENT OF ANGLE AFTER TOTAL HIP-REPLACEMENT [J].
ACKLAND, MK ;
BOURNE, WB ;
UHTHOFF, HK .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1986, 68 (03) :409-413
[2]
[Anonymous], 1979, CLIN ORTHOP RELAT RE
[3]
COMPLICATIONS RELATED TO MODULARITY OF TOTAL HIP COMPONENTS [J].
BARRACK, RL ;
BURKE, DW ;
COOK, SD ;
SKINNER, HB ;
HARRIS, WH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1993, 75 (05) :688-692
[4]
Berman AT, 1996, ORTHOPEDICS, V19, P971
[5]
DISSOCIATION OF ACETABULAR COMPONENTS AFTER TOTAL HIP-ARTHROPLASTY - REPORT OF 4 CASES [J].
BRIEN, WW ;
SALVATI, EA ;
WRIGHT, TM ;
NELSON, CL ;
HUNGERFORD, DS ;
GILLIAM, DL .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (10) :1548-1550
[6]
Cameron H U, 1993, Orthop Rev, V22, P1160
[7]
An analysis of Food and Drug Administration medical device reports relating to total joint components [J].
Castro, FP ;
Chimento, G ;
Munn, BG ;
Levy, RS ;
Timon, S ;
Barrack, RL .
JOURNAL OF ARTHROPLASTY, 1997, 12 (07) :765-771
[8]
Claus B, 1993, Acta Orthop Belg, V59 Suppl 1, P307
[9]
Primary hybrid total hip replacement, performed with insertion of the acetabular component without cement and a precoat femoral component with cement - An average ten-year follow-up study [J].
Clohisy, JC ;
Harris, WH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1999, 81A (02) :247-255
[10]
The Harris-Galante porous-coated acetabular component with screw fixation - An average ten-year follow-up study [J].
Clohisy, JC ;
Harris, WH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1999, 81A (01) :66-73