Quality assurance for the geometric accuracy of cone-beam CT guidance in radiation therapy

被引:56
作者
Bissonnette, Jean-Pierre [1 ]
Moseley, Doug [1 ]
White, Elizabeth [1 ]
Sharpe, Michael [1 ]
Purdie, Tom [1 ]
Jaffray, David A. [1 ]
机构
[1] Princess Margaret Hosp, Radiat Med Program, Toronto, ON M5G 2M9, Canada
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2008年 / 71卷 / 01期
关键词
image guidance; geometric accuracy; cone-beam CT;
D O I
10.1016/j.ijrobp.2007.06.086
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The introduction of volumetric X-ray image-guided radiotherapy systems allows improved management of geometric variations in patient setup and internal organ motion. As these systems become a routine clinical modality, we propose a daily quality assurance (QA) program for cone-beam computed tomography (CBCT) integrated with a linear accelerator. The image-guided system used in this work combines a linear accelerator with conventional X-ray tube and an amorphous silicon flat-panel detector mounted orthogonally from the accelerator central beam axis. This article focuses on daily QA protocols germane to geometric accuracy of the CBCT systems and proposes tolerance levels on the basis of more than 3 years of experience with seven CBCT systems used in our clinic. Monthly geometric calibration tests demonstrate the long-term stability of the flex movements, which are reproducible within +/- 0.5 mm (95% confidence interval). The daily QA procedure demonstrates that, for rigid phantoms, the accuracy of the image-guided process can be within 1 mm on average, with a 99% confidence interval of +/- 2 mm. (C) 2008 Elsevier Inc.
引用
收藏
页码:S57 / S61
页数:5
相关论文
共 14 条
[1]  
[Anonymous], 2002, 74 AAPM
[2]   Accurate technique for complete geometric calibration of cone-beam computed tomography systems [J].
Cho, YB ;
Moseley, DJ ;
Siewerdsen, JH ;
Jaffray, DA .
MEDICAL PHYSICS, 2005, 32 (04) :968-983
[3]   Flat-panel cone-beam computed tomography for image-guided radiation therapy [J].
Jaffray, DA ;
Siewerdsen, JH ;
Wong, JW ;
Martinez, AA .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (05) :1337-1349
[4]   A radiographic and tomographic imaging system integrated into a medical linear accelerator for localization of bone and soft-tissue targets [J].
Jaffray, DA ;
Drake, DG ;
Moreau, M ;
Martinez, AA ;
Wong, JW .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 45 (03) :773-789
[5]  
JAFFRAY DA, 2005, PHYS MODERN TECHNOLO, V2, P259
[6]   Performance and stability of the varian cone-beam CT system [J].
Kim, G. ;
Pawlicki, T. ;
Luxton, G. .
MEDICAL PHYSICS, 2006, 33 (06) :2004-2004
[7]   COMPREHENSIVE QA FOR RADIATION ONCOLOGY - REPORT OF AAPM RADIATION-THERAPY COMMITTEE TASK-GROUP-40 [J].
KUTCHER, GJ ;
COIA, L ;
GILLIN, M ;
HANSON, WF ;
LEIBEL, S ;
MORTON, RJ ;
PALTA, JR ;
PURDY, JA ;
REINSTEIN, LE ;
SVENSSON, GK ;
WELLER, M ;
WINGFIELD, L .
MEDICAL PHYSICS, 1994, 21 (04) :581-618
[8]  
Lin P-JP, 1993, 39 AAPM, P1
[9]   Quantifying the geometric accuracy of the on board imager over a one-year period [J].
Lovelock, D. ;
LoSasso, T. ;
Ali, I. ;
Amols, H. ;
Ling, C. ;
Yamada, Y. ;
Pham, H. ;
Munro, P. .
MEDICAL PHYSICS, 2006, 33 (06) :2000-2001
[10]   A SYSTEM FOR STEREOTACTIC RADIOSURGERY WITH A LINEAR-ACCELERATOR [J].
LUTZ, W ;
WINSTON, KR ;
MALEKI, N .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1988, 14 (02) :373-381