Metal ion levels after metal-on-metal proximal femoral replacements - A 30-year follow-up

被引:71
作者
Dunstan, E [1 ]
Sanghrajka, AP [1 ]
Tilley, S [1 ]
Unwin, P [1 ]
Blunn, G [1 ]
Cannon, SR [1 ]
Briggs, TWR [1 ]
机构
[1] Royal Natl Orthopaed Hosp, Stanmore HA7 4LP, Middx, England
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 2005年 / 87B卷 / 05期
关键词
D O I
10.1302/0301-620X.87B5.15384
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Metal-on-metal hip bearings are being implanted into younger patients. The consequence of elevated levels of potentially carcinogenic metal ions is therefore a cause for concern. We have determined the levels of cobalt (Co), chromium (Cr), titanium (Ti) and vanadium (Va) in the urine and whole blood of patients who had had metal-on-metal and metal-on-polyethylene articulations in situ for more than 30 years. We compared these with each other and with the levels for a control group of subjects. We found significantly elevated levels of whole blood Ti, Va and urinary Cr in all arthroplasty groups. The whole blood and urine levels of Co were grossly elevated, by a factor of 50 and 300 times respectively in patients with loose metal-on-metal articulations when compared with the control group. Stable metal-on-metal articulations showed much lower levels. Elevated levels of whole blood or urinary Co may be useful in identifying metal-on-metal articulations which are loose.
引用
收藏
页码:628 / 631
页数:4
相关论文
共 22 条
[1]   The wear pattern in metal-on-metal hip prostheses [J].
Anissian, HL ;
Stark, A ;
Good, V ;
Dahlstrand, H ;
Clarke, IC .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 2001, 58 (06) :673-678
[2]  
ARNSTUTA HC, 1996, CLIN ORTHOP S, V329, P11
[3]   Elevated serum cobalt with metal-on-metal articulating surfaces [J].
Brodner, W ;
Bitzan, P ;
Meisinger, V ;
Kaider, A ;
GottsaunerWolf, F ;
Kotz, R .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1997, 79B (02) :316-321
[4]   WIDESPREAD DISSEMINATION OF METAL DEBRIS FROM IMPLANTS [J].
CASE, CP ;
LANGKAMER, VG ;
JAMES, C ;
PALMER, MR ;
KEMP, AJ ;
HEAP, PF ;
SOLOMON, L .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (05) :701-712
[5]   Levels of metal ions after small- and large-diameter metal-on-metal hip arthroplasty [J].
Clarke, MT ;
Lee, PTH ;
Arora, A ;
Villar, RN .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2003, 85B (06) :913-917
[6]   CONCENTRATION OF WEAR PRODUCTS IN HAIR, BLOOD, AND URINE AFTER TOTAL HIP REPLACEMENT [J].
COLEMAN, RF ;
HERRINGTON, J ;
SCALES, JT .
BMJ-BRITISH MEDICAL JOURNAL, 1973, 1 (5852) :527-529
[7]   ENDOPROSTHETIC REPLACEMENT OF THE PROXIMAL FEMUR AND ACETABULUM - A SURVIVAL ANALYSIS [J].
DOBBS, HS ;
SCALES, JT ;
WILSON, JN ;
KEMP, HBS ;
BURROWS, HJ ;
SNEATH, RS .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1981, 63 (02) :219-224
[8]   THE INCIDENCE OF CANCER FOLLOWING TOTAL HIP-REPLACEMENT [J].
GILLESPIE, WJ ;
FRAMPTON, CMA ;
HENDERSON, RJ ;
RYAN, PM .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1988, 70 (04) :539-542
[9]   MALIGNANCY AND JOINT REPLACEMENT [J].
GOODFELLOW, J .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (05) :645-645
[10]  
HARRIS WH, 1995, CLIN ORTHOP RELAT R, P46