Uncertainty budgets for calibration of radiation thermometers below the silver point

被引:69
作者
Saunders, P. [1 ]
Fischer, J. [2 ]
Sadli, M. [3 ]
Battuello, M. [4 ]
Park, C. W. [5 ]
Yuan, Z. [6 ]
Yoon, H. [7 ]
Li, W. [8 ]
van der Ham, E. [9 ]
Sakuma, F. [10 ]
Ishii, J. [10 ]
Ballico, M. [11 ]
Machin, G. [12 ]
Fox, N. [12 ]
Hollandt, J. [2 ]
Matveyev, M. [13 ]
Bloembergen, P. [10 ]
Ugur, S. [14 ]
机构
[1] Ind Res Ltd, Measurement Stand Lab New Zealand MSL, Lower Hutt, New Zealand
[2] Phys Tech Bundesanstalt, D-3300 Braunschweig, Germany
[3] CNAM, INM, LNE, La Plaine St Denis, France
[4] Ist Nazl Ric Metrol INRiM, Turin, Italy
[5] KRISS, Taejon, South Korea
[6] NIM, Beijing, Peoples R China
[7] Natl Inst Stand & Technol, Gaithersburg, MD 20899 USA
[8] SPRING, NMC, Stand Prod & Innovat Board, Singapore, Singapore
[9] Nederlands Meetinst Swinden Lab BV NMi VSL, Delft, Netherlands
[10] AIST, NMIJ, Tsukuba, Ibaraki, Japan
[11] NMIA, Lindfield, NSW, Australia
[12] NPL, Teddington, Middx, England
[13] Rostekhregulirovaniye Russia VNIIM, DI Mendeleyev Inst Metrol, St Petersburg, Russia
[14] TEKNOYAD, Gebze, Kocaeli, Turkey
关键词
calibration; radiation thermometry; uncertainty;
D O I
10.1007/s10765-008-0385-1
中图分类号
O414.1 [热力学];
学科分类号
摘要
Below the freezing point of silver, radiation thermometers are generally calibrated by implementing the multi-point interpolation method using blackbody measurements at three or more calibration points, rather than the ITS-90 extrapolation technique. The interpolation method eliminates the need to measure the spectral responsivity and provides greater accuracy at the longer wavelengths required below the silver point. This article identifies all the sources of uncertainty associated with the interpolation method, in particular, those related to the reference blackbody temperatures (either variable-temperature or fixed-point blackbodies) and to the measured thermometer signals at these points. Estimates are given of the 'normal' and 'best' uncertainties currently achievable. A model of the thermometer response is used to propagate all the uncertainties at the reference points and provide a total uncertainty at any temperature within the calibration range. The multi-point method has the effect of constraining the total uncertainty over this range, unlike the ITS-90 technique for which the uncertainties propagate as T-2. This article is a joint effort of the working group on radiation thermometry of the Consultative Committee for Thermometry (CCT), summarizing the knowledge and experience of all experts in this field.
引用
收藏
页码:1066 / 1083
页数:18
相关论文
共 63 条
[1]   Impurity dependence of the aluminium point [J].
Ancsin, J .
METROLOGIA, 2003, 40 (02) :36-41
[2]  
[Anonymous], P TEMPEMKO 1999 7 IN
[3]  
[Anonymous], P TEMPEMKO 1996 6 IN
[4]  
[Anonymous], 1997, P 6 INT S TEMP THERM
[5]  
[Anonymous], P TEMPMEKO 1999 7 IN
[6]   A simple technique for measuring the infrared emissivity of black-body radiators [J].
Ballico, M .
METROLOGIA, 2000, 37 (04) :295-300
[7]  
Ballico M, 2002, TEMPMEKO 2001: 8TH INTERNATIONAL SYMPOSIUM ON TEMPERATURE AND THERMAL MEASUREMENT IN INDUSTRY AND SCIENCE, VOL 1 & 2, PROCEEDINGS, P233
[8]  
BALLICO M, COMMUNICATION
[9]  
Ballico M., 2004, P TEMPMEKO 2004 9 IN, P841
[10]   Modelling of the effective emissivity of a graphite tube black body [J].
Ballico, MJ .
METROLOGIA, 1996, 32 (04) :259-265