Determination of mercury in hair by square-wave anodic stripping voltammetry at a rotating gold disk electrode after microwave digestion

被引:17
作者
East, GA [1 ]
Marinho, EP [1 ]
机构
[1] Univ Brasilia, Inst Quim, BR-70910900 Brasilia, DF, Brazil
关键词
mercury determination; hair analysis; microwave digestion; anodic stripping voltammetry; rotating gold electrode;
D O I
10.1385/BTER:103:3:261
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 [生物化学与分子生物学]; 081704 [应用化学];
摘要
A simple and reliable method for the determination of mercury in hair on a rotating gold disk electrode using subtractive anodic stripping voltammetry without removal of oxygen is reported. Voltammetric and microwave parameters were optimized to obtain the best analytical results. Parameters such as supporting electrolyte concentration, influence of chloride in the Hg peak, deposition potential, scan rate, accumulation time, rotation rate, square-wave amplitude, and electrode conditioning were studied. Pressurized microwave-assisted digestion of hair, suitable for the accurate voltammetric determination of Hg, was evaluated using six acid mixtures and several time-power programs. Under the optimized conditions, no interference by copper, cadmium, lead, nickel, manganese, iron, or zinc was found at concentrations corresponding to their occurrence in normal hair. A calibration plot between 6,67 and 46,69 μ g/L was linear, with r(2) better than 0.999. The detection limit for a deposition time of 60 s at 254g was calculated as 1.92 nM (3σ). Validation of the method was demonstrated with the use of a certified reference sample of hair. Eight real samples of hair (four unexposed children and four exposed persons) were also analyzed.
引用
收藏
页码:261 / 276
页数:16
相关论文
共 26 条
[1]
Determination of transition metals in human hair by high-performance liquid chromatography using sodium hexadecane-sulfonate coated columns [J].
Almeida, AA ;
Jun, X ;
Lima, JLFC .
TALANTA, 1999, 50 (02) :253-259
[2]
DETERMINATION OF TRACE MERCURY(II) IN 0.1 M-PERCHLORIC ACID BY DIFFERENTIAL PULSE STRIPPING VOLTAMMETRY AT A ROTATING GOLD DISK ELECTRODE [J].
ANDREWS, RW ;
LAROCHELLE, JH ;
JOHNSON, DC .
ANALYTICAL CHEMISTRY, 1976, 48 (01) :212-214
[3]
MERCURY CONTAMINATION IN THE BRAZILIAN AMAZON - ENVIRONMENTAL, AND OCCUPATIONAL ASPECTS [J].
BARBOSA, AC ;
BOISCHIO, AA ;
EAST, GA ;
FERRARI, I ;
GONCALVES, A ;
SILVA, PRM ;
DACRUZ, TME .
WATER AIR AND SOIL POLLUTION, 1995, 80 (1-4) :109-121
[4]
Trace determination of mercury by anodic stripping voltammetry at the rotating gold electrode [J].
Bonfil, Y ;
Brand, M ;
Kirowa-Eisner, E .
ANALYTICA CHIMICA ACTA, 2000, 424 (01) :65-76
[5]
THE ASSESSMENT OF REFERENCE VALUES FOR ELEMENTS IN HUMAN BIOLOGICAL TISSUES AND FLUIDS - A SYSTEMATIC REVIEW [J].
CAROLI, S ;
ALIMONTI, A ;
CONI, E ;
PETRUCCI, F ;
SENOFONTE, O ;
VIOLANTE, N .
CRITICAL REVIEWS IN ANALYTICAL CHEMISTRY, 1994, 24 (5-6) :363-398
[6]
Easy-to-make Ag/AgCl reference electrode [J].
East, GA ;
del Valle, MA .
JOURNAL OF CHEMICAL EDUCATION, 2000, 77 (01) :97-97
[7]
Stripping voltammetric determination of mercury at modified solid electrodes - Determination of mercury traces using PDC/Au(III) modified electrodes [J].
Faller, C ;
Stozhko, NY ;
Henze, G ;
Brainina, KZ .
ANALYTICA CHIMICA ACTA, 1999, 396 (2-3) :195-202
[8]
GIL EP, 1994, ANAL CHIM ACTA, V293, P55
[9]
RAPID HIGH-PERFORMANCE ANALYTICAL PROCEDURE WITH SIMULTANEOUS VOLTAMMETRIC DETERMINATION OF TOXIC TRACE-METALS IN URINE [J].
GOLIMOWSKI, J ;
VALENTA, P ;
STOEPPLER, M ;
NURNBERG, HW .
TALANTA, 1979, 26 (08) :649-656
[10]
ANODIC-STRIPPING SEMI-DIFFERENTIAL ELECTROANALYSIS WITH THIN MERCURY FILM ELECTRODE FORMED INSITU [J].
GOTO, M ;
IKENOYA, K ;
ISHII, D .
ANALYTICAL CHEMISTRY, 1979, 51 (01) :110-115