Clinical comparative study of cardiopulmonary bypass with Nikkiso and BioMedicus centrifugal pumps

被引:8
作者
Ohtsubo, S
Tayama, E
Short, D
Noon, GP
Nose, Y
机构
[1] Department of Surgery, Baylor College of Medicine, Houston, TX
[2] Department of Surgery, Baylor College of Medicine, Houston, TX 77030, One Baylor Plaza
关键词
cardiopulmonary bypass; Nikkiso centrifugal pump; BioMedicus centrifugal pump;
D O I
10.1111/j.1525-1594.1996.tb04510.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The Nikkiso centrifugal pump was evaluated in elective adult open heard surgery in comparison with the BioMedicus pump. Ten patients using the Nikkiso pump (Group N), and 10 patients using the BioMedicus pump (Group B) were examined for (or to determine) hematobiologic parameters and patient outcome data as well as pump controllability. During cardiopulmonary bypass (CPB), both pumps maintained systemic perfusion satisfactorily without any mechanical adverse event. Rotation speed of the Nikkiso centrifugal pump (3,580 +/- 100 rpm) was significantly higher than that of the BioMedicus pump (3,170 +/- 100 rpm; p < 0.05) whereas changes in free plasma hemoglobin, platelet count, blood urea nitrogen, and creatinine levels showed no significant differences between the two groups. Urine output in Group N for 30 min after the initiation of CPB (7.10 +/- 1.50 ml/kg/h) was significantly higher than that in Group B (3.23 +/- 0.46 ml/kg/h; p < 0.05). Patient outcome data were similar in both groups, such as duration of intensive care unit stay, hospital stay, postoperative intubation time, amount of postoperative bleeding, and amount of blood transfused. These equivalent results with the BioMedicus pump suggested that the Nikkiso pump can be used in open heart surgery as a reliable and atraumatic CPB pump.
引用
收藏
页码:715 / 720
页数:6
相关论文
共 8 条
[1]  
[Anonymous], J EXTRA CORPOR TECHN
[2]  
Araki K., 1994, Japanese Journal of Artificial Organs, V23, P898
[3]  
BURKI T, 1992, VASCULAR SURG, V26, P123
[4]  
Jakob H G, 1991, ASAIO Trans, V37, pM487
[5]   BEWARE CENTRIFUGAL PUMPS - NOT A ONE-WAY STREET, BUT A POTENTIALLY DANGEROUS SIPHON [J].
KOLFF, J ;
MCCLURKEN, JB ;
ALPERN, JB .
ANNALS OF THORACIC SURGERY, 1990, 50 (03) :512-512
[6]   INITIAL CLINICAL-EXPERIENCE WITH THE BAYLOR-NIKKISO CENTRIFUGAL PUMP [J].
OHTSUBO, S ;
NAITO, K ;
MATSUURA, M ;
KAWAHITO, K ;
SHIMONO, T ;
MAKINOUCHI, K ;
TASAI, K ;
OHARA, Y ;
DAMM, G ;
GLUECK, J ;
RASKIN, S ;
TAKATANI, S ;
BENKOWSKI, R ;
SHORT, DH ;
SCHINEN, SA ;
NOON, GP ;
NOSE, Y .
ARTIFICIAL ORGANS, 1995, 19 (07) :769-773
[7]   CLINICAL-EXPERIENCE WITH NIKKISO CENTRIFUGAL PUMPS FOR EXTRACORPOREAL-CIRCULATION [J].
ONODA, K ;
KONDO, C ;
MIZUMOTO, T ;
KUSAGAWA, H ;
KATAYAMA, Y ;
HAYASHI, T ;
KOMADA, T ;
HIRANO, R ;
MIYAMURA, T ;
TANAKA, J ;
YASUDA, F ;
TANAKA, K ;
SHIMPO, H ;
YADA, I ;
YUASA, H ;
AIZAWA, T ;
KUSAGAWA, M .
ARTIFICIAL ORGANS, 1994, 18 (09) :706-710
[8]   THE EFFECTS OF PRESSURE AND FLOW ON HEMOLYSIS CAUSED BY BIO-MEDICUS CENTRIFUGAL PUMPS AND ROLLER-PUMPS - GUIDELINES FOR CHOOSING A BLOOD PUMP [J].
TAMARI, Y ;
LEESENSIBA, K ;
LEONARD, EF ;
PARNELL, V ;
TORTOLANI, AJ .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1993, 106 (06) :997-1007