Intrathoracic impedance monitoring in patients with heart failure - Correlation with fluid status and feasibility of early warning preceding hospitalization

被引:501
作者
Yu, CM [1 ]
Wang, L
Chau, E
Chan, RHW
Kong, SL
Tang, MO
Christensen, J
Stadler, RW
Lau, CP
机构
[1] Chinese Univ Hong Kong, Prince Wales Hosp, Dept Med & Therapeut, Div Cardiol, Shatin, Hong Kong, Peoples R China
[2] Medtronic Inc, Minneapolis, MN USA
[3] Grantham Hosp, Hong Kong, Hong Kong, Peoples R China
[4] Univ Hong Kong, Queen Mary Hosp, Dept Med, Div Cardiol, Hong Kong, Hong Kong, Peoples R China
关键词
heart failure; pulmonary wedge pressure; pacing;
D O I
10.1161/CIRCULATIONAHA.104.492207
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background - Patients with heart failure are frequently hospitalized for fluid overload. A reliable method for chronic monitoring of fluid status is therefore desirable. We evaluated an implantable system capable of measuring intrathoracic impedance to identify potential fluid overload before heart failure hospitalization and to determine the correlation between intrathoracic impedance and standard measures of fluid status during hospitalization. Methods and Results - Thirty-three patients with NYHA class III and IV heart failure were implanted with a special pacemaker in the left pectoral region and a defibrillation lead in the right ventricle. Intrathoracic impedance was regularly measured and recorded between the lead and the pacemaker case. During hospitalizations, pulmonary capillary wedge pressure and fluid status were monitored. Ten patients were hospitalized for fluid overload 25 times over 20.7 +/- 8.4 months. Intrathoracic impedance decreased before each admission by an average of 12.3 +/- 5.3% ( P < 0.001) over an average of 18.3 +/- 10.1 days. Impedance reduction began 15.3 +/- 10.6 days ( P < 0.001) before the onset of worsening symptoms. There was an inverse correlation between intrathoracic impedance and pulmonary capillary wedge pressure ( r = - 0.61, P < 0.001) and between intrathoracic impedance and net fluid loss ( r = - 0.70, P < 0.001) during hospitalization. Automated detection of impedance decreases was 76.9% sensitive in detecting hospitalization for fluid overload, with 1.5 false-positive ( threshold crossing without hospitalization) detections per patient-year of follow-up. Conclusions - Intrathoracic impedance is inversely correlated with pulmonary capillary wedge pressure and fluid balance and decreased before the onset of patient symptoms and before hospital admission for fluid overload. Regular monitoring of impedance may provide early warning of impending decompensation and diagnostic information for titration of medication.
引用
收藏
页码:841 / 848
页数:8
相关论文
共 29 条
[1]   Cardiac resynchronization in chronic heart failure [J].
Abraham, WT ;
Fisher, WG ;
Smith, AL ;
Delurgio, DB ;
Leon, AR ;
Loh, E ;
Kocovic, DZ ;
Packer, M ;
Clavell, AL ;
Hayes, DL ;
Ellestad, M ;
Messenger, J ;
Trupp, RJ ;
Underwood, J ;
Pickering, F ;
Truex, C ;
McAtee, P .
NEW ENGLAND JOURNAL OF MEDICINE, 2002, 346 (24) :1845-1853
[2]   Continuous autonomic assessment in patients with symptomatic heart failure - Prognostic value of heart rate variability measured by an implanted cardiac resynchronization device [J].
Adamson, PB ;
Smith, AL ;
Abraham, WT ;
Kleckner, KJ ;
Stadler, RW ;
Shih, A ;
Rhodes, MM .
CIRCULATION, 2004, 110 (16) :2389-2394
[3]  
*AM HEART ASS, 2003, HEART DIS STROK STAT
[4]   NONINVASIVE MEASUREMENT OF INTRATHORACIC FLUIDS [J].
BAKER, LE ;
DENNISTON, JC .
CHEST, 1974, 65 (04) :S35-S37
[5]  
Bennett S J, 1998, Am J Crit Care, V7, P168
[6]   Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure [J].
Bristow, MR ;
Saxon, LA ;
Boehmer, J ;
Krueger, S ;
Kass, DA ;
De Marco, T ;
Carson, P ;
DiCarlo, L ;
DeMets, D ;
White, BG ;
DeVries, DW ;
Feldman, AM .
NEW ENGLAND JOURNAL OF MEDICINE, 2004, 350 (21) :2140-2150
[7]   Transthoracic monitoring of the impedance of the right lung in patients with cardiogenic pulmonary edema [J].
Charach, G ;
Rabinovich, P ;
Grosskopf, I ;
Weintraub, M .
CRITICAL CARE MEDICINE, 2001, 29 (06) :1137-1144
[8]  
COMBS W, 1994, THORACIC IMPEDANCE M, P240
[9]   PARAMETERS OF TRANSTHORACIC ELECTRICAL CONDUCTION [J].
COOLEY, WL .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1970, 170 (02) :702-&
[10]   Older adults' symptoms and their duration before hospitalization for heart failure [J].
Friedman, MM .
HEART & LUNG, 1997, 26 (03) :169-176