Selective preconcentration and ion chromatography of trace lead(II) in environmental samples using a porous graphitic carbon column

被引:7
作者
O'Connell, MP [1 ]
Treacy, J [1 ]
Merly, C [1 ]
Smith, CMM [1 ]
Glennon, JD [1 ]
机构
[1] Univ Coll Cork, Dept Chem, Div Analyt Chem, Cork, Ireland
关键词
porous graphitic carbon; ion chromatography; metal preconcentration; lead(II);
D O I
10.1080/00032719908542607
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Highly selective ion analysis of sub-ppm Pb(II) on a porous graphitic carbon (PGC) column, is described using ion chromatography with acidified aqueous mobile phases, post-column detection using PAR and spectrophotometric detection. Sensitivity is enhanced by on-line preconcentration of the metal prior to the chromatography using one millilitre injection volumes applied to a calix[4]arene-tetrahydroxamate based chelating precolumn. The method is applied to water samples collected from a river subjected to an industrial effluent discharge and results obtained compare favourably with those obtained using flame atomic absorption spectroscopy.
引用
收藏
页码:185 / 192
页数:8
相关论文
共 8 条
[1]   BIOCHELATION CARTRIDGE FOR THE SOLID-PHASE EXTRACTION OF TRACE-METALS [J].
GLENNON, JD ;
SRIJARANAI, S .
ANALYST, 1990, 115 (05) :627-630
[2]   SEPARATION OF ANIONIC AND CATIONIC COMPOUNDS OF BIOMEDICAL INTEREST BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY ON POROUS GRAPHITIC CARBON [J].
GU, GH ;
LIM, CK .
JOURNAL OF CHROMATOGRAPHY, 1990, 515 :183-192
[3]   SOLID-PHASE EXTRACTION OF METAL-IONS USING IMMOBILIZED CHELATING CALIXARENE TETRAHYDROXAMATES [J].
HUTCHINSON, S ;
KEARNEY, GA ;
HORNE, E ;
LYNCH, B ;
GLENNON, JD ;
MCKERVEY, MA ;
HARRIS, SJ .
ANALYTICA CHIMICA ACTA, 1994, 291 (03) :269-275
[4]   Surface modification of porous graphite for ion exchange chromatography [J].
Knox, JH ;
Wan, QH .
CHROMATOGRAPHIA, 1996, 42 (1-2) :83-88
[5]   Selective ion chromatography of metals on porous graphitic carbon [J].
Merly, C ;
Lynch, B ;
Ross, P ;
Glennon, JD .
JOURNAL OF CHROMATOGRAPHY A, 1998, 804 (1-2) :187-192
[6]   Determination of calcium and magnesium in sea-water using a dynamically coated porous graphitic carbon column with a selective metallochromic ligand as a component of the mobile phase [J].
Paull, B ;
Fagan, PA ;
Haddad, PR .
ANALYTICAL COMMUNICATIONS, 1996, 33 (06) :193-196
[7]  
ROSS P, 1997, ADV CHROMATOGRAPHY A, V37, P73
[8]  
RYAN N, 1992, ANAL PROC, V29, P21