Practical strategies for the clinical implementation of matrix inversion tomosynthesis (MITS)

被引:25
作者
Godfrey, DJ [1 ]
Rader, A [1 ]
Dobbins, JT [1 ]
机构
[1] Duke Univ, Dept Biomed Engn, Durham, NC 27710 USA
来源
MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2 | 2003年 / 5030卷
关键词
tomosynthesis; linear tomography; limited-angle tomography; tomographic reconstruction; tomographic blur; digital radiography; flat-panel detector; matrix inversion; chest radiography; image retention;
D O I
10.1117/12.480352
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
Digital tomosynthesis is a method that enables the retroactive reconstruction of arbitrary tomographic planes in an object from a finite series of digital projection radiographs, acquired with limited, angle tube movement. Conventional tomosynthesis suffers from the presence of blurring artifacts, created by objects located outside of each reconstructed plane. Matrix inversion tomosynthesis (MITS) utilizes known acquisition geometry to solve directly for the unwanted out-of-plane blur artifacts, thus enabling their removal. This paper examines practical strategies for the implementation of MITS in a clinical setting, on a flat-panel fast-readout detector, with the aim of minimizing procedure time and image reconstruction artifacts concurrently. Topics include a comparison of continuous vs. incremental tube motion, the presence of reconstruction artifacts due to error in computing the x-ray tube location, the effect of scrubbing the detector between projections to reduce image retention, and a method for accounting for data that gets projected off the detector. We conclude that MITS is robust enough to be clinically applicable, even under less-than-ideal conditions. Rapid-image acquisition with continuous tube movement and no detector scrubbing is clinically desirable. for MITS imaging of the chest, where patient motion is a concern. Knowledge of the source-detector geometry can be satisfactorily,determined via either a lead fiducial marker placed on the patient, or a tube motion device with sufficient precision and accuracy. Extrapolation of data at the top and bottom of projection images provides excellent amelioration of image truncation artifacts.
引用
收藏
页码:379 / 390
页数:12
相关论文
共 25 条
[1]
Resolution improvement in linear tomosynthesis with an adapted 3D regularization scheme [J].
Bleuet, P ;
Guillemaud, W ;
Magnin, IE .
MEDICAL IMAGING 2002: PHYSICS OF MEDICAL IMAGING, 2002, 4682 :117-125
[2]
Technical and clinical results of an experimental Flat Dynamic (digital) X-ray image Detector (FDXD) system with real-time corrections [J].
Bruijns, TJC ;
Alving, PL ;
Baker, EL ;
Bury, R ;
Cowen, AR ;
Jung, N ;
Luijendijk, HA ;
Meulenbrugge, HJ ;
Stouten, HJ .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :33-44
[3]
SELF-MASKING SUBTRACTION TOMOSYNTHESIS [J].
CHAKRABORTY, DP ;
YESTER, MV ;
BARNES, GT ;
LAKSHMINARAYANAN, AV .
RADIOLOGY, 1984, 150 (01) :225-229
[4]
ITERATIVE 3-DIMENSIONAL IMAGE-RECONSTRUCTION FROM TOMOGRAPHIC PROJECTIONS [J].
COLSHER, JG .
COMPUTER GRAPHICS AND IMAGE PROCESSING, 1977, 6 (06) :513-537
[5]
des Plantes BGZ, 1932, ACTA RADIOL, V13, P182
[6]
Dobbins III James T., 1990, US Patent, Patent No. [4903204, US4903204 A]
[7]
REDUCTION OF LINEAR BLURRING IN TOMOGRAPHY [J].
EDHOLM, PR ;
QUIDING, L .
RADIOLOGY, 1969, 92 (05) :1115-&
[8]
Optimization of matrix inverse tomosynthesis [J].
Godfrey, DJ ;
Warp, RJ ;
Dobbins, JT .
MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 :696-704
[10]
Dynamic X-ray imaging system based on an amorphous silicon thin-film array [J].
Jung, N ;
Alving, PL ;
Busse, F ;
Conrads, N ;
Meulenbrugge, HM ;
Rutten, W ;
Schiebel, U ;
Weibrecht, M ;
Wieczorek, H .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :396-407