Improvement in Lesion Detection with Whole-Body Oncologic Time-of-Flight PET

被引:159
作者
El Fakhri, Georges [2 ,3 ]
Surti, Suleman
Trott, Cathryn M. [2 ,3 ]
Scheuermann, Joshua
Karp, Joel S. [1 ]
机构
[1] Univ Penn, Sch Med, Dept Radiol, Philadelphia, PA 19104 USA
[2] Harvard Univ, Sch Med, Dept Radiol, Div Nucl Med & Mol Imaging, Boston, MA 02115 USA
[3] Massachusetts Gen Hosp, Boston, MA 02114 USA
基金
美国国家卫生研究院;
关键词
WB-(18)FDG; TOF-PET; lesion detection; OBSERVER PERFORMANCE; RECONSTRUCTION; ATTENUATION; TOMOGRAPHY; SCANNER; BENEFIT; IMPACT;
D O I
10.2967/jnumed.110.080382
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Time-of-flight (TOF) PET has great potential in whole-body oncologic applications, and recent work has demonstrated qualitatively in patient studies the improvement that can be achieved in lesion visibility. The aim of this work was to objectively quantify the improvement in lesion detectability that can be achieved in lung and liver lesions with whole-body F-18-FDG TOF PET in a cohort of 100 patients as a function of body mass index, lesion location and contrast, and scanning time. Methods: One hundred patients with BMIs ranging from 16 to 45 were included in this study. Artificial 1-cm spheric lesions were imaged separately in air at variable locations of each patient's lung and liver, appropriately attenuated, and incorporated in the patient list-mode data with 4 different lesion-to-background contrast ranges. The fused studies with artificial lesion present or absent were reconstructed using a list-mode unrelaxed ordered-subsets expectation maximization with chronologically ordered subsets and a gaussian TOF kernel for TOF reconstruction. Conditions were compared on the basis of performance of a 3-channel Hotelling observer signal-to-noise ratio in detecting the presence of a sphere of unknown size on an anatomic background while modeling observer noise. Results: TOF PET yielded an improvement in lesion detection performance (3-channel Hotelling observer signal-to-noise ratio) over non-TOF PET of 8.3% in the liver and 15.1% in the lungs. The improvement in all lesions was 20.3%, 12.0%, 9.2%, and 7.5% for mean contrast values of 2.0:1, 3.2:1, 4.4:1, and 5.7:1, respectively. Furthermore, this improvement was 9.8% in patients with a BMI of less than 30 and 11.1% in patients with a BMI of 30 or more. Performance plateaued faster as a function of number of iterations with TOF than non-TOF. Conclusion: Over all contrasts and body mass indexes, oncologic TOF PET yielded a significant improvement in lesion detection that was greater for lower lesion contrasts. This improvement was achieved without compromising other aspects of PET imaging.
引用
收藏
页码:347 / 353
页数:7
相关论文
共 19 条
[1]
Practical issues and methodology in assessment of image quality using model observers [J].
Abbey, CK ;
Barrett, HH ;
Eckstein, MP .
PHYSICS OF MEDICAL IMAGING - MEDICAL IMAGING 1997, 1997, 3032 :182-194
[2]
Human- and model-observer performance in ramp-spectrum noise: effects of regularization and object variability [J].
Abbey, CK ;
Barrett, HH .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2001, 18 (03) :473-488
[3]
[Anonymous], 1998, NIH PUBL
[4]
BUDINGER TF, 1983, J NUCL MED, V24, P73
[5]
First experimental results of time-of-flight reconstruction on an LSO PET scanner [J].
Conti, M ;
Bendriem, B ;
Casey, M ;
Chen, M ;
Kehren, F ;
Michel, C ;
Panin, V .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (19) :4507-4526
[6]
Impact of acquisition geometry, image processing, and patient size on lesion detection in whole-body 18F-FDG PET [J].
El Fakhri, Georges ;
Santos, Paula A. ;
Badawi, Ramsey D. ;
Holdsworth, Clay H. ;
Van den Abbeele, Annick D. ;
Kijewski, Marie Foley .
JOURNAL OF NUCLEAR MEDICINE, 2007, 48 (12) :1951-1960
[7]
Farquhar TH, 1999, J NUCL MED, V40, P2043
[8]
HOTELLING TRACE CRITERION AND ITS CORRELATION WITH HUMAN-OBSERVER PERFORMANCE [J].
FIETE, RD ;
BARRETT, HH ;
SMITH, WE ;
MYERS, KJ .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1987, 4 (05) :945-953
[9]
Experimental Comparison of Lesion Detectability for Four Fully-3D PET Reconstruction Schemes [J].
Kadrmas, Dan J. ;
Casey, Michael E. ;
Black, Noel F. ;
Hamill, James J. ;
Panin, Vladimir Y. ;
Conti, Maurizio .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2009, 28 (04) :523-534
[10]
Benefit of time-of-flight in PET: Experimental and clinical results [J].
Karp, Joel S. ;
Surti, Sulernan ;
Daube-Witherspoon, Margaret E. ;
Muehllehner, Gerd .
JOURNAL OF NUCLEAR MEDICINE, 2008, 49 (03) :462-470