Equilibrium in-fibre standardisation technique for solid-phase microextraction

被引:115
作者
Wang, YX [1 ]
O'Reilly, J [1 ]
Chen, Y [1 ]
Pawliszyn, J [1 ]
机构
[1] Univ Waterloo, Dept Chem, Waterloo, ON N2L 3G1, Canada
关键词
solid-phase microextraction; calibration; internal standardisation; automation;
D O I
10.1016/j.chroma.2004.12.084
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This note describes a fundamental investigation into solid-phase microextraction (SPME) using a standard loaded into the fibre coating as a means of internal standardisation for the analysis of samples contained in vials. The loading of reproducible amounts of standards into a non-porous SPME fiber was investigated. It was found that spiking low milligram quantities of standards such as benzene, toluene, ethylbenzene, xylenes (BTEX) and/or naphthalene into a few grams of pump oil sealed in a 20 mL vial provided an excellent standard generator. A single solution allowed over a hundred standard loadings with a reproducibility of < 4% R.S.D. When a fiber, loaded with the standard(s) was introduced into a sample vial, extraction of analytes into the fiber and desorption of the standard(s) into the sample matrix occur simultaneously. Quantification was then based on the equilibrium distribution of the standards and the analytes between the fibre coating and the sample matrix in the vial. A comparison of equilibration profiles obtained using traditional internal standardisation and the in-fibre approach generally showed the same equilibration behaviour. The developed method was successfully used to correct for matrix effects in the BTEX analysis of a wine sample. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:13 / 17
页数:5
相关论文
共 8 条
[1]   Solid phase microextraction for quantitative analysis in nonequilibrium situations [J].
Ai, J .
ANALYTICAL CHEMISTRY, 1997, 69 (06) :1230-1236
[2]   Headspace solid phase microextraction. Dynamics and quantitative analysis before reaching a partition equilibrium [J].
Ai, J .
ANALYTICAL CHEMISTRY, 1997, 69 (16) :3260-3266
[3]   Kinetics and the on-site application of standards in a solid-phase microextration fiber [J].
Chen, Y ;
Pawliszyn, J .
ANALYTICAL CHEMISTRY, 2004, 76 (19) :5807-5815
[4]   Standards in the extraction phase, a new approach to calibration of microextraction processes [J].
Chen, Y ;
O'Reilly, J ;
Wang, YX ;
Pawliszyn, J .
ANALYST, 2004, 129 (08) :702-703
[5]   LIMITS OF THE INTERNAL STANDARD TECHNIQUE IN CHROMATOGRAPHY [J].
HAEFELFINGER, P .
JOURNAL OF CHROMATOGRAPHY, 1981, 218 (1-3) :73-81
[6]  
Pawliszyn J., 1997, SOLID PHASE MICROEXT
[7]  
SCHEPPERSWERCIN.SA, 1999, SOLID PHASE MICROEXT
[8]  
US EPA (United States Environmental Protection Agency), 2003, 5021A US EPA