Dual lumen transducer probes for real-time 3-D interventional cardiac ultrasound

被引:25
作者
Lee, W
Idriss, SF
Wolf, PD
Smith, SW
机构
[1] Duke Univ, Dept Biomed Engn, Durham, NC 27708 USA
[2] Duke Univ, Med Ctr, Dept Pediat, Durham, NC 27710 USA
关键词
2-D array; volumetric ultrasound imaging; intracardiac echocardiography; forward-viewing catheter;
D O I
10.1016/S0301-5629(03)00952-9
中图分类号
O42 [声学];
学科分类号
070206 [声学]; 082403 [水声工程];
摘要
We have developed dual lumen probes incorporating a forward-viewing matrix array transducer with an integrated working lumen for delivery of tools in real-time 3-D (RT3-D) interventional echocardiography. The probes are of 14 Fr and 22 Fr sizes, with 112 channel 2-D arrays operating at 5 MHz. We obtained images of cardiac anatomy and simultaneous interventional device delivery with an in vivo sheep model, including: manipulation of a 0.36-mm diameter guidewire into the coronary sinus, guidance of a transseptal puncture using a 1.2-mm diameter Brockenbrough needle, and guidance of a right ventricular biopsy using 3 Fr biopsy forceps. We have also incorporated the 22 Fr probe within a 6-mm surgical trocar to obtain apical four-chamber ultrasound (US) scans from a subcostal position. Combining the imaging catheter with a working lumen in a single device may simplify cardiac interventional procedures by allowing clinicians to easily visualize cardiac structures and simultaneously direct interventional tools in a RT3-D image. (C) 2003 World Federation for Ultrasound in Medicine Biology.
引用
收藏
页码:1297 / 1304
页数:8
相关论文
共 23 条
[1]
Abraham WT, 2002, J AM COLL CARDIOL, V39, p171A
[2]
Abraham WT, 2002, J AM COLL CARDIOL, V39, p159A
[3]
Real-time three-dimensional dobutamine stress echocardiography in assessment of ischemia: Comparison with two-dimensional dobutamine stress echocardiography [J].
Ahmad, M ;
Xie, TR ;
McCulloch, M ;
Abreo, G ;
Runge, M .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (05) :1303-1309
[4]
Intracardiac and intraluminal echocardiography:: Indications and standard approaches [J].
Bartel, T ;
Müller, S ;
Caspari, G ;
Erbel, R .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2002, 28 (08) :997-1003
[5]
Blanc JJ, 1997, CIRCULATION, V96, P3273
[6]
Intracardiac Doppler hemodynamics and flow: New vector, phased-array ultrasound-tipped catheter [J].
Bruce, CJ ;
Packer, DL ;
Seward, JB .
AMERICAN JOURNAL OF CARDIOLOGY, 1999, 83 (10) :1509-+
[7]
Quantitative assessment of left ventricular perfusion defects using real-time three-dimensional myocardial contrast echocardiography [J].
Camarano, G ;
Jones, M ;
Freidlin, RZ ;
Panza, JA .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2002, 15 (03) :206-213
[8]
CAVES PK, 1974, LANCET, V1, P821
[9]
RADIOFREQUENCY CATHETER ABLATION GUIDED BY INTRACARDIAC ECHOCARDIOGRAPHY [J].
CHU, E ;
FITZPATRICK, AP ;
CHIN, MC ;
SUDHIR, K ;
YOCK, PG ;
LESH, MD .
CIRCULATION, 1994, 89 (03) :1301-1305
[10]
High-density flexible interconnect for two-dimensional ultrasound arrays [J].
Fiering, JO ;
Hultman, P ;
Lee, W ;
Light, ED ;
Smith, SW .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2000, 47 (03) :764-770