Automated on-line separation-preconcentration system for platinum determination by electrothermal atomic absorption spectrometry

被引:44
作者
Ojeda, CB [1 ]
Rojas, FS [1 ]
Pavón, JMC [1 ]
de Torres, AG [1 ]
机构
[1] Univ Malaga, Fac Sci, Dept Analyt Chem, E-29071 Malaga, Spain
关键词
platinum; electrothermal atomic absorption spectrometry; preconcentration; microcolumn; flow injection;
D O I
10.1016/S0003-2670(03)01014-6
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A flow injection on-line sorption preconcentration system for the electrothermal atomic absorption spectrometric determination of platinum has been developed. The method employs on-line preconcentration of platinum on a column packed with silica gel functionalised with 1,5-bis(di-2-pyridyl)methylene thiocarbohydrazide (DPTH-gel) placed in the autosampler arm. The modification of the autosampler in the tubing line and circuit allowed either the flow of the sample through the column or the operation of the autosampler in the normal mode, where microlitres of 2 M HNO3, which acts as the elution agent, pass through the microcolumn eluting Pt(IV), which is directly deposited in the graphite tube as a drop of a precisely defined volume. An enrichment factor of 41.7 and a 3sigma detection limit of 0.8 ng ml(-1) along with a sampling frequency of 29 h(-1) are achieved with 60 s preconcentration time at a sample flow rate of 2.4 ml min(-1). The relative standard deviation is 1.6% for 4 ng ml(-1) (n = 11). The method has been applied to the determination of platinum in catalyst, vegetation, soil and water samples. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:97 / 103
页数:7
相关论文
共 19 条
[1]   PLATINUM SALT SENSITIVITY IN REFINERY WORKERS - INCIDENCE AND EFFECTS OF SMOKING AND EXPOSURE [J].
CALVERLEY, AE ;
REES, D ;
DOWDESWELL, RJ ;
LINNETT, PJ ;
KIELKOWSKI, D .
OCCUPATIONAL AND ENVIRONMENTAL MEDICINE, 1995, 52 (10) :661-666
[2]   ONLINE PRECONCENTRATION AND DETERMINATION OF TRACE PLATINUM BY FLOW-INJECTION ATOMIC-ABSORPTION SPECTROMETRY [J].
CANTARERO, A ;
GOMEZ, MM ;
CAMARA, C ;
PALACIOS, MA .
ANALYTICA CHIMICA ACTA, 1994, 296 (02) :205-211
[3]   ONLINE PRECONCENTRATION AND SEPARATION OF PALLADIUM, PLATINUM AND IRIDIUM USING ALPHA-AMINO PYRIDINE RESIN WITH FLAME ATOMIC-ABSORPTION SPECTROMETRY [J].
DI, P ;
DAVEY, DE .
TALANTA, 1995, 42 (05) :685-692
[4]  
Fang Z., 1995, FLOW INJECTION ATOMI
[5]   FUNDAMENTAL AND PRACTICAL CONSIDERATIONS IN THE DESIGN OF ONLINE COLUMN PRECONCENTRATION FOR FLOW-INJECTION ATOMIC SPECTROMETRIC SYSTEMS [J].
FANG, ZL ;
XU, SK ;
ZHANG, SC .
ANALYTICA CHIMICA ACTA, 1987, 200 (01) :35-49
[6]   DETERMINATION OF PALLADIUM AND PLATINUM IN FRESH-WATERS BY INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY AND ACTIVATED-CHARCOAL PRECONCENTRATION [J].
HALL, GEM ;
PELCHAT, JC .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (08) :1059-1065
[7]  
Hidalgo MM, 1996, FRESEN J ANAL CHEM, V354, P420
[8]  
JAN B, 1986, CLIN PHARMACOL THER, V39, P136
[9]   PRECONCENTRATION OF PALLADIUM, PLATINUM AND RHODIUM BY ONLINE SORBENT EXTRACTION FOR GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY AND INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
LEE, ML ;
TOLG, G ;
BEINROHR, E ;
TSCHOPEL, P .
ANALYTICA CHIMICA ACTA, 1993, 272 (02) :193-203
[10]   Evaluation of a flow injection system combined with an inductively coupled plasma mass spectrometer with thermospray nebulization for the determination of trace levels of platinum [J].
Parent, M ;
Vanhoe, H ;
Moens, L ;
Dams, R .
ANALYTICA CHIMICA ACTA, 1996, 320 (01) :1-10