Picomolar detection limits with current-polarized Pb2+ ion-selective membranes

被引:130
作者
Pergel, E
Gyurcsányi, RE
Tóth, K
Lindner, E
机构
[1] Univ Tennessee, Hlth Sci Ctr, Herff Coll Engn, Memphis, TN 38152 USA
[2] Univ Memphis, Joint Grad Program Biomed Engn, Memphis, TN 38152 USA
[3] Budapest Univ Technol & Econ, Inst Gen & Analyt Chem, H-1111 Budapest, Hungary
[4] Hungarian Acad Sci, Res Grp Tech Analyt Chem, H-1111 Budapest, Hungary
关键词
D O I
10.1021/ac010094a
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Minor ion fluxes across ion-selective membranes bias submicromolar activity measurements with conventional ion-selective electrodes. When ion fluxes are balanced, the lower limit of detection is expected to be dramatically improved. As proof of principle, the flux of lead ions across an ETH 5435 ionophore-based lead-selective membrane was gradually compensated by applying a few nanoamperes of galvanostatic current. When the opposite ion fluxes were matched, and the undesirable leaching of primary ions was eliminated, Nernstian response down to 3 x 10(-12) M was achieved.
引用
收藏
页码:4249 / 4253
页数:5
相关论文
共 30 条
[1]   EXCHANGE CURRENT DETERMINATION AND TIME-DEPENDENT EFFECTS FOR MODEL LIQUID MEMBRANE ION-SELECTIVE ELECTRODES [J].
ARMSTRONG, RD ;
COVINGTON, AK ;
EVANS, GP ;
HANDYSIDE, T .
ELECTROCHIMICA ACTA, 1984, 29 (08) :1127-1131
[2]   Potentiometry at trace levels [J].
Bakker, E ;
Pretsch, E .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2001, 20 (01) :11-19
[3]   Selectivity of potentiometric ion sensors [J].
Bakker, E ;
Pretsch, E ;
Bühlmann, P .
ANALYTICAL CHEMISTRY, 2000, 72 (06) :1127-1133
[4]   Carrier-based ion-selective electrodes and bulk optodes. 1. General characteristics [J].
Bakker, E ;
Buhlmann, P ;
Pretsch, E .
CHEMICAL REVIEWS, 1997, 97 (08) :3083-3132
[5]  
BUCK RP, 1994, ANAL PROC, V31, P301
[6]  
CAMMANN K, 1985, ION SELECTIVE ELECTR, P35
[7]   Influence of key parameters on the lower detection limit and response function of solvent polymeric membrane ion-selective electrodes [J].
Ceresa, A ;
Sokalski, T ;
Pretsch, E .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2001, 501 (1-2) :70-76
[8]   PVC MATRIX MEMBRANE ION-SELECTIVE ELECTRODES - CONSTRUCTION AND LABORATORY EXPERIMENTS [J].
CRAGGS, A ;
MOODY, GJ ;
THOMAS, JDR .
JOURNAL OF CHEMICAL EDUCATION, 1974, 51 (08) :541-544
[9]  
*EPA, 1999, 815F899010 EPA
[10]   Direct evidence of ionic fluxes across ion selective membranes:: A scanning electrochemical microscopic and potentiometric study [J].
Gyurcsányi, RE ;
Pergel, E ;
Nagy, R ;
Kapui, I ;
Lan, BTT ;
Tóth, K ;
Bitter, I ;
Lindner, E .
ANALYTICAL CHEMISTRY, 2001, 73 (09) :2104-2111