Direct evidence of ionic fluxes across ion selective membranes:: A scanning electrochemical microscopic and potentiometric study

被引:106
作者
Gyurcsányi, RE
Pergel, E
Nagy, R
Kapui, I
Lan, BTT
Tóth, K
Bitter, I
Lindner, E [1 ]
机构
[1] Univ Memphis, Joint Grad Program Biomed Engn, Memphis, TN 38152 USA
[2] Univ Tennessee, Hlth Sci Ctr, Herff Coll Engn, Knoxville, TN 37996 USA
[3] Budapest Univ Technol & Econ, Inst Gen & Analyt Chem, Hungarian Acad Sci, Res Grp Tech Analyt Chem, H-1111 Budapest, Hungary
[4] Budapest Univ Technol & Econ, Inst Gen & Analyt Chem, H-1111 Budapest, Hungary
[5] Budapest Univ Technol & Econ, Inst Organ Chem Technol, H-1111 Budapest, Hungary
关键词
D O I
10.1021/ac000922k
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Scanning electrochemical microscopy (SECM) supplemented with potentiometric measurements was used to follow the time-dependent buildup of a steady-state diffusion layer at the aqueous-phase boundary of lead ion-selective electrodes (ISEs). Differential pulse voltammetry is adapted to SECM for probing the local concentration profiles at the sample side of solvent polymeric membranes. Major factors affecting the membrane transport-related surface concentrations were identified from SECM data and the potentiometric transients obtained under different experimental conditions (inner filling solution composition, membrane thickness, surface pretreatment). The amperometrically determined surface. concentrations correlated well with the lower detection limits of the lead ion-selective electrodes.
引用
收藏
页码:2104 / 2111
页数:8
相关论文
共 62 条
[1]   Mapping concentration profiles within the diffusion layer of an electrode Part III. Steady-state and time-dependent profiles via amperometric measurements with an ultramicroelectrode probe [J].
Amatore, C ;
Szunerits, S ;
Thouin, L ;
Warkocz, JS .
ELECTROCHEMISTRY COMMUNICATIONS, 2000, 2 (05) :353-358
[2]   LIPOPHILIC SALTS AS MEMBRANE ADDITIVES AND THEIR INFLUENCE ON THE PROPERTIES OF MACROELECTRODES AND MICROELECTRODES BASED ON NEUTRAL CARRIERS [J].
AMMANN, D ;
PRETSCH, E ;
SIMON, W ;
LINDNER, E ;
BEZEGH, A ;
PUNGOR, E .
ANALYTICA CHIMICA ACTA, 1985, 171 (MAY) :119-129
[3]   Permeation of ammonia across bilayer lipid membranes studied by ammonium ion selective microelectrodes [J].
Antonenko, YN ;
Pohl, P ;
Denisov, GA .
BIOPHYSICAL JOURNAL, 1997, 72 (05) :2187-2195
[4]   WEAK ACID TRANSPORT ACROSS BILAYER-LIPID MEMBRANE IN THE PRESENCE OF BUFFERS - THEORETICAL AND EXPERIMENTAL PH PROFILES IN THE UNSTIRRED LAYERS [J].
ANTONENKO, YN ;
DENISOV, GA ;
POHL, P .
BIOPHYSICAL JOURNAL, 1993, 64 (06) :1701-1710
[5]   PROPERTIES OF PVC BASED MEMBRANES USED IN ION-SELECTIVE ELECTRODES [J].
ARMSTRONG, RD ;
HORVAI, G .
ELECTROCHIMICA ACTA, 1990, 35 (01) :1-7
[6]   Ion sensors: current limits and new trends [J].
Bakker, E ;
Diamond, D ;
Lewenstam, A ;
Pretsch, E .
ANALYTICA CHIMICA ACTA, 1999, 393 (1-3) :11-18
[7]   DETECTION LIMIT OF ION-SELECTIVE BULK OPTODES AND CORRESPONDING ELECTRODES [J].
BAKKER, E ;
WILLER, M ;
PRETSCH, E .
ANALYTICA CHIMICA ACTA, 1993, 282 (02) :265-271
[8]   Carrier-based ion-selective electrodes and bulk optodes. 1. General characteristics [J].
Bakker, E ;
Buhlmann, P ;
Pretsch, E .
CHEMICAL REVIEWS, 1997, 97 (08) :3083-3132
[9]   SCANNING ELECTROCHEMICAL MICROSCOPY - INTRODUCTION AND PRINCIPLES [J].
BARD, AJ ;
FAN, FRF ;
KWAK, J ;
LEV, O .
ANALYTICAL CHEMISTRY, 1989, 61 (02) :132-138
[10]   CHEMICAL IMAGING OF SURFACES WITH THE SCANNING ELECTROCHEMICAL MICROSCOPE [J].
BARD, AJ ;
FAN, FRF ;
PIERCE, DT ;
UNWIN, PR ;
WIPF, DO ;
ZHOU, FM .
SCIENCE, 1991, 254 (5028) :68-74