Noninvasive high frequency oscillatory ventilation through nasal prongs: bench evaluation of efficacy and mechanics

被引:71
作者
De Luca, Daniele [1 ,2 ]
Carnielli, Virgilio P. [1 ,2 ,3 ,4 ]
Conti, Giorgio [5 ]
Piastra, Marco [5 ]
机构
[1] Polytech Univ Marche, Div Neonatol, G Salesi Women & Children Hosp, Ancona, Italy
[2] Polytech Univ Marche, Inst Mother & Child Hlth, G Salesi Women & Children Hosp, Ancona, Italy
[3] Inst Child Hlth, London, England
[4] Great Ormond St Hosp Sick Children, London, England
[5] Univ Cattolica Sacro Cuore, Pediat Intens Care Unit, Dept Emergency & Intens Care, I-00168 Rome, Italy
关键词
Bench study; HFOV; Noninvasive ventilation; Preterm neonate; GAS-TRANSPORT; PERFORMANCE; IMPEDANCE;
D O I
10.1007/s00134-010-2054-7
中图分类号
R4 [临床医学];
学科分类号
100218 [急诊医学];
摘要
Noninvasive high frequency oscillatory ventilation through nasal prongs (nHFOV) has been proposed as a new respiratory support in neonatology. We studied the effect of ventilation parameters and nasal prongs on nHFOV efficacy and mechanics. Customized sealed circuits connecting a SM3100A oscillator to a neonatal lung model were developed to evaluate the effect of applying HFOV via two different sized nasal prongs on delivered tidal volume and pressure. Measurements were made across a range of frequencies and pressures; amplitude was set to obtain visible lung oscillation. Volume delivered by peak-to-peak oscillation, ventilation, and pressure significantly differed among the interfaces, being higher for large cannulae and the control circuit (p < 0.0001). The interposition of a large or small nasal prong reduced volume to 56 and 26%, ventilation to 32 and 9%, and mean pressure to 83 and 79%, respectively, of the values measured for the direct connection of the oscillator to the test lung. Volume and ventilation were inversely related to frequency, which was particularly evident with larger diameter circuits due to higher delivered tidal volume (R (2) > 0.9). Increasing ventilation was associated with larger tidal volume and nasal prong diameter (adjusted R (2) = 0.97). nHFOV using common nasal prongs is technically possible. Efficiency of tidal volume delivery is significantly affected by prong diameter.
引用
收藏
页码:2094 / 2100
页数:7
相关论文
共 25 条
[1]
GAS-EXCHANGE IN HEALTHY RABBITS DURING HIGH-FREQUENCY OSCILLATORY VENTILATION [J].
BOYNTON, BR ;
HAMMOND, MD ;
FREDBERG, JJ ;
BUCKLEY, BG ;
VILLANUEVA, D ;
FRANTZ, ID .
JOURNAL OF APPLIED PHYSIOLOGY, 1989, 66 (03) :1343-1351
[2]
BOYNTON BR, 1994, NEW THERAPIES NEONAT, P218
[3]
Cameron AC, 1997, J ECONOMETRICS, V77, P329
[4]
Should nasal high-frequency ventilation be used in preterm infants? [J].
Carlo, Waldemar A. .
ACTA PAEDIATRICA, 2008, 97 (11) :1484-1485
[5]
MECHANISMS OF GAS-TRANSPORT DURING VENTILATION BY HIGH-FREQUENCY OSCILLATION [J].
CHANG, HK .
JOURNAL OF APPLIED PHYSIOLOGY, 1984, 56 (03) :553-563
[6]
Nasal high-frequency ventilation for premature infants [J].
Colaizy, Tarah T. ;
Younis, Usama M. M. ;
Bell, Edward F. ;
Klein, Jonathan M. .
ACTA PAEDIATRICA, 2008, 97 (11) :1518-1522
[7]
In vitro comparison of nasal continuous positive airway pressure devices for neonates [J].
De Paoli, AG ;
Morley, CJ ;
Davis, PG ;
Lau, R ;
Hingeley, E .
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2002, 87 (01) :F42-F45
[8]
High-frequency oscillatory ventilation for acute respiratory distress syndrome in adult patients [J].
Derdak, S .
CRITICAL CARE MEDICINE, 2003, 31 (04) :S317-S323
[9]
Influence of upper airway shunt on total respiratory impedance in infants [J].
Desager, KN ;
Cauberghs, M ;
Naudts, J ;
Van de Woestijne, KP .
JOURNAL OF APPLIED PHYSIOLOGY, 1999, 87 (03) :902-909
[10]
FACTORS INFLUENCING MECHANICAL PERFORMANCE OF NEONATAL HIGH-FREQUENCY VENTILATORS [J].
FREDBERG, JJ ;
GLASS, GM ;
BOYNTON, BR ;
FRANTZ, ID .
JOURNAL OF APPLIED PHYSIOLOGY, 1987, 62 (06) :2485-2490