Integrated microfluidic device for solid-phase extraction coupled to micellar electrokinetic chromatography separation

被引:78
作者
Ramsey, JD
Collins, GE
机构
[1] USN, Res Lab, Div Chem, Washington, DC 20375 USA
[2] Geocenters Inc, Arlington, VA 22201 USA
关键词
D O I
10.1021/ac0507789
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An integrated microdevice was utilized for the autonomous coupling of solid-phase extraction (SPE) to micellar electrokinetic chromatography (MEKC). Porous plugs of polymethacrylate polymer (similar to 200 mu m in length) were fabricated by ultraviolet irradiation in microchannels. Microcolumns of hydrophobic beads packed against the polymethacrylate plugs were utilized for the quantitative extraction of rhodamine B, yielding preconcentration factors over 200 for a 90-s extraction. The calculated detection limit for this dye was 60 M. A sample of coumarin dyes were concentrated by SPE, eluted in a nonaqueous solvent from a separate on-chip reservoir, and injected by a gated valve onto a separate column for MEKC analysis. Using the integrated device, a completely automated sequence of extraction, elution, injection, separation, and detection were performed in less than 5 min. Observed separation efficiencies were high, with plate heights below 2 mu m. The analysis was at least 3 times faster than semiautomated, conventional, solid-phase extraction, while requiring no user intervention. The design, fabrication, and autonomous operation of the device is discussed.
引用
收藏
页码:6664 / 6670
页数:7
相关论文
共 45 条
[1]   Micromachined flow-through filter-chamber for chemical reactions on beads [J].
Andersson, H ;
van der Wijngaart, W ;
Enoksson, P ;
Stemme, G .
SENSORS AND ACTUATORS B-CHEMICAL, 2000, 67 (1-2) :203-208
[2]  
Bergkvist J, 2002, PROTEOMICS, V2, P422, DOI 10.1002/1615-9861(200204)2:4<422::AID-PROT422>3.0.CO
[3]  
2-1
[4]   Sample filtration, concentration, and separation integrated on microfluidic devices [J].
Broyles, BS ;
Jacobson, SC ;
Ramsey, JM .
ANALYTICAL CHEMISTRY, 2003, 75 (11) :2761-2767
[5]   An integrated fritless column for on-chip capillary electrochromatography with conventional stationary phases [J].
Ceriotti, L ;
de Rooij, NF ;
Verpoorte, E .
ANALYTICAL CHEMISTRY, 2002, 74 (03) :639-647
[6]   Macroporous photopolymer frits for capillary electrochromatography [J].
Chen, JR ;
Dulay, MT ;
Zare, RN ;
Svec, F ;
Peters, E .
ANALYTICAL CHEMISTRY, 2000, 72 (06) :1224-1227
[7]   Developments toward a complete micro-total analysis system for Duchenne muscular dystrophy diagnosis [J].
Ferrance, JP ;
Wu, QR ;
Giordano, B ;
Hernandez, C ;
Kwok, Y ;
Snow, K ;
Thibodeau, S ;
Landers, JP .
ANALYTICA CHIMICA ACTA, 2003, 500 (1-2) :223-236
[8]   Integrated capillary electrophoresis devices with an efficient postcolumn reactor in planar quartz and glass chips [J].
Fluri, K ;
Fitzpatrick, G ;
Chiem, N ;
Harrison, DJ .
ANALYTICAL CHEMISTRY, 1996, 68 (23) :4285-4290
[9]   Integrated microfluidic system enabling protein digestion, peptide separation, and protein identification [J].
Gao, J ;
Xu, JD ;
Locascio, LE ;
Lee, CS .
ANALYTICAL CHEMISTRY, 2001, 73 (11) :2648-2655
[10]   Integrated microchip-device for the digestion, separation and postcolumn labeling of proteins and peptides [J].
Gottschlich, N ;
Culbertson, CT ;
McKnight, TE ;
Jacobson, SC ;
Ramsey, JM .
JOURNAL OF CHROMATOGRAPHY B, 2000, 745 (01) :243-249