Effect of delayed scanning of storage phosphor plates

被引:24
作者
Akdeniz, BG [1 ]
Gröndahl, HG
Kose, T
机构
[1] Ege Univ, Dishekimligi Fak, Oral Diagnoz & Rad AD, TR-35100 Izmir, Turkey
[2] Univ Gothenburg, Sahlgrenska Acad, Fac Odontol, Gothenburg, Sweden
[3] Ege Univ, Sch Engn, Dept Comp Engn, TR-35100 Izmir, Turkey
来源
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY | 2005年 / 99卷 / 05期
关键词
D O I
10.1016/j.tripleo.2004.10.021
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective. To test longevity of image quality in storage phosphor plates (SPPs) at various exposure settings, storage conditions, and delays in scanning. Study design. Fifteen Digora plates were exposed from 0.08 to 0.20 seconds and scanned immediately, 10, 30, and 60 minutes, and 24 hours after exposure. Plates were stored both in daylight and in a light-tight box. Mean gray values (MGVs) were compared using 2 x 5 x 5 factorial ANOVA. Interaction between variables was tested using Bonferroni/Dunn multiple comparisons test. Results. MGVs decreased with increase in exposure but increased with the scan delay. Only MGVs of plates scanned within 10 minutes after exposure were not significantly different from the ones scanned immediately (P > .05). MGVs increased with scan delay for all exposure times no matter how the plates were stored (P < .05). Conclusion. Based on the time delays examined, it is recommended to scan the Digora SPP no later than 10 minutes after exposure. Longer periods may cause loss of quality.
引用
收藏
页码:603 / 607
页数:5
相关论文
共 14 条
[1]   Radiographic image enhancement. Part I: spatial domain techniques [J].
Analoui, M .
DENTOMAXILLOFACIAL RADIOLOGY, 2001, 30 (01) :1-9
[2]   On the dynamic range of different X-ray photon detectors in intra-oral radiography. A comparison of image quality in film, charge-coupled device and storage phosphor systems [J].
Borg, E ;
Grondahl, HG .
DENTOMAXILLOFACIAL RADIOLOGY, 1996, 25 (02) :82-88
[3]   Subjective image quality of solid-state and photostimulable phosphor systems for digital intra-oral radiography [J].
Borg, E ;
Attaelmanan, A ;
Gröndahl, HG .
DENTOMAXILLOFACIAL RADIOLOGY, 2000, 29 (02) :70-75
[4]   Observers' use of image enhancement in assessing caries in radiographs taken by four intra-oral digital systems [J].
Gotfredsen, E ;
Wenzel, A ;
Grondahl, HG .
DENTOMAXILLOFACIAL RADIOLOGY, 1996, 25 (01) :34-38
[5]  
Hildebolt C F, 2000, Dent Clin North Am, V44, P273
[6]   Accuracy of caries detection with four storage phosphor systems and E-speed radiographs [J].
Hintze, H ;
Wenzel, A ;
Frydenberg, M .
DENTOMAXILLOFACIAL RADIOLOGY, 2002, 31 (03) :170-175
[7]   Comparison of a photostimulable phosphor system with film for dental radiology [J].
Huda, W ;
Rill, LN ;
Benn, DK ;
Pettigrew, JC .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1997, 83 (06) :725-731
[8]   Comparison of three intra-oral storage phosphor systems using subjective image quality [J].
Kitagawa, H ;
Farman, AG ;
Scheetz, JP ;
Brown, WP ;
Lewis, J ;
Benefiel, M ;
Kuroyanagi, K .
DENTOMAXILLOFACIAL RADIOLOGY, 2000, 29 (05) :272-276
[9]   Analysis of digital images acquired using different phosphor storage plates (PSPs) subjected to varying reading times and storage conditions [J].
Martins, MGBQ ;
Neto, FH ;
Whaites, EJ .
DENTOMAXILLOFACIAL RADIOLOGY, 2003, 32 (03) :186-190
[10]   Detection of approximal caries with a storage phosphor system. A comparison of enhanced digital images with dental X-ray film [J].
Moystad, A ;
Svanaes, DB ;
Risnes, S ;
Larheim, TA ;
Grondahl, HG .
DENTOMAXILLOFACIAL RADIOLOGY, 1996, 25 (04) :202-206