The highly sensitive detection of catecholamines using a microfluidic device integrated with an enzyme-modified pre-reactor for interferent elimination and an interdigitated array electrode

被引:28
作者
Hayashi, K
Iwasaki, Y
Kurita, R
Sunagawa, K
Niwa, O
Tate, A
机构
[1] NTT Corp, Microsyst Integrat Labs, Kanagawa 2430198, Japan
[2] NTT Corp, Adv Technol Corp, Kanagawa 2430198, Japan
[3] Natl Cardiovasc Ctr, Osaka 5658565, Japan
关键词
catecholamines; enzyme-modified pre-reactor; interdigitated array electrode; continuous monitoring;
D O I
10.1016/j.jelechem.2005.01.033
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We developed a microfluidic device for the continuous monitoring of catecholamine levels and studied the performance of two components of the device, namely an enzyme-modified pre-reactor and an interdigitated array (IDA) electrode. The enzyme-modified pre-reactor and IDA electrode were integrated on a chip. The integrated pre-reactor with a 3D structure, consisted of rectangular-shaped micropillars immobilized with ascorbate oxidase, and was used for the enzymatic removal of the electroactive interferent, L-ascorbic acid (AA). It exhibited elimination efficiencies of 99.7% and 98% for AA concentrations of 10 and 100 mu M, respectively. We used a carbon-based IDA electrode (2 mu m line and space) fabricated in a microflow channel for the highly sensitive detection of CAs. Redox cycling at the IDA electrode and an increase in the mass transport to the electrode contributed to the highly sensitive detection of dopamine (DA) and adrenaline (AD). The detection limits of the device for DA and AD standard solution in the absence of AA were 108 and 440 pM, respectively. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:215 / 222
页数:8
相关论文
共 38 条
[1]   ELECTROCHEMICAL RESPONSE AT MICROARRAY ELECTRODES IN FLOWING STREAMS AND DETERMINATION OF CATECHOLAMINES [J].
AOKI, A ;
MATSUE, T ;
UCHIDA, I .
ANALYTICAL CHEMISTRY, 1990, 62 (20) :2206-2210
[2]  
CAHILL PS, 1996, ANAL CHEM, V68, P3160
[3]  
CHEN Z, IN PRESS ELECTROANAL
[4]   Simultaneous determination of catecholamines in rat brain tissue by high-performance liquid chromatography [J].
Chi, JD ;
Odontiadis, J ;
Franklin, M .
JOURNAL OF CHROMATOGRAPHY B, 1999, 731 (02) :361-367
[5]   PLASMA NOREPINEPHRINE AS A GUIDE TO PROGNOSIS IN PATIENTS WITH CHRONIC CONGESTIVE HEART-FAILURE [J].
COHN, JN ;
LEVINE, TB ;
OLIVARI, MT ;
GARBERG, V ;
LURA, D ;
FRANCIS, GS ;
SIMON, AB ;
RECTOR, T .
NEW ENGLAND JOURNAL OF MEDICINE, 1984, 311 (13) :819-823
[6]  
Collins A, 2001, ELECTROANAL, V13, P425, DOI 10.1002/1521-4109(200104)13:6<425::AID-ELAN425>3.0.CO
[7]  
2-5
[8]   Dissociation of dopamine release in the nucleus accumbens from intracranial self-stimulation [J].
Garris, PA ;
Kilpatrick, M ;
Bunin, MA ;
Michael, D ;
Walker, QD ;
Wightman, RM .
NATURE, 1999, 398 (6722) :67-69
[9]   Continuous separations with microfabricated electrophoresis-electrochemical array detection [J].
Gavin, PF ;
Ewing, AG .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1996, 118 (37) :8932-8936
[10]   ELECTROCHEMICAL STUDIES OF OXIDATION PATHWAYS OF CATECHOLAMINES [J].
HAWLEY, MD ;
TATAWAWA.SV ;
PIEKARSK.S ;
ADAMS, RN .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1967, 89 (02) :447-&