Simultaneous MR/PET imaging of the human brain: Feasibility study

被引:365
作者
Schlemmer, Heinz-Peter W. [1 ]
Pichler, Bernd J. [1 ]
Schmand, Matthias [2 ]
Burbar, Ziad [2 ]
Michel, Christian [2 ]
Ladebeck, Ralf [3 ]
Jattke, Kirstin [3 ]
Townsend, David [4 ]
Nahmias, Claude [4 ]
Jacob, Pradeep K. [5 ]
Heiss, Wolf-Dieter [6 ]
Claussen, Claus D. [1 ]
机构
[1] Univ Tubingen Hosp, Dept Diagnost Radiol, D-72076 Tubingen, Germany
[2] Siemens Med Solut, Dept Mol Imaging, Knoxville, TN USA
[3] Siemens Med Solut, Dept Magnet Resonance, Erlangen, Germany
[4] Univ Tennessee, Grad Sch Med, Mol Imaging & Translat Res Program, Knoxville, TN USA
[5] Univ Tennessee, Med Ctr, Dept Radiol, Knoxville, TN USA
[6] Max Planck Inst Neurol Res, Cologne, Germany
关键词
D O I
10.1148/radiol.2483071927
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
The purpose of this study was to apply a magnetic resonance (MR) imaging-compatible positron emission tomographic (PET) detector technology for simultaneous MR/PET imaging of the human brain and skull base. The PET detector ring consists of lutetium oxyorthosilicate (LSO) scintillation crystals in combination with avalanche photodiodes (APDs) mounted in a clinical 3-T MR imager with use of the birdcage transmit/receive head coil. Following phantom studies, two patients were simultaneously examined by using fluorine 18 fluorodeoxyglucose (FDG) PET and MR imaging and spectroscopy. MR/PET data enabled accurate coregistration of morphologic and multifunctional information. Simultaneous MR/PET imaging is feasible in humans, opening up new possibilities for the emerging field of molecular imaging. (C) RSNA, 2008.
引用
收藏
页码:1028 / 1035
页数:8
相关论文
共 18 条
[1]
Beyer T, 2000, J NUCL MED, V41, P1369
[2]
Byars LG, 2005, IEEE NUCL SCI CONF R, P2622
[3]
Comtat C, 2004, IEEE NUCL SCI CONF R, P3492
[4]
Goerres GW, 2004, J NUCL MED, V45, p66S
[5]
Heiss WD, 2004, J NUCL MED, V45, P1811
[6]
3-D PHANTOM TO SIMULATE CEREBRAL BLOOD-FLOW AND METABOLIC IMAGES FOR PET [J].
HOFFMAN, EJ ;
CUTLER, PD ;
DIGBY, WM ;
MAZZIOTTA, JC .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1990, 37 (02) :616-620
[7]
HOFMANN M, 2008, J NUCL MED IN PRESS
[8]
PET/MR images acquired with a compact MR-compatiable PET detector in a 7-T magnet [J].
Judenhofer, Martin S. ;
Catana, Ciprian ;
Swann, Brian K. ;
Siegel, Stefan B. ;
Jung, Wulf-Ingo ;
Nutt, Robert E. ;
Cherry, Simon R. ;
Claussen, Claus D. ;
Pichler, Bernd J. .
RADIOLOGY, 2007, 244 (03) :807-814
[9]
CERIUM-DOPED LUTETIUM OXYORTHOSILICATE - A FAST, EFFICIENT NEW SCINTILLATOR [J].
MELCHER, CL ;
SCHWEITZER, JS .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1992, 39 (04) :502-505
[10]
Prospective comparison of the impact on treatment decisions of whole-body magnetic resonance imaging and computed tomography in patients with metastatic malignant melanoma [J].
Müller-Horvat, C ;
Radny, P ;
Eigentler, TK ;
Schäfer, J ;
Pfannenberg, C ;
Horger, M ;
Khorchidi, S ;
Nägele, T ;
Garbe, C ;
Claussen, CD ;
Schlemmer, HP .
EUROPEAN JOURNAL OF CANCER, 2006, 42 (03) :342-350