Nickel content dependence of electrochemical behavior of carbohydrates on a titanium-nickel alloy electrode and its application to a liquid chromatography detector

被引:25
作者
Morita, M
Niwa, O
Tou, S
Watanabe, N
机构
[1] NTT, Basic Res Labs, Atsugi, Kanagawa 2430198, Japan
[2] BAS Japan Ltd, Tokyo 1318790, Japan
[3] Univ Tokyo, Fac Engn, Dept Appl Chem, Tokyo 1138656, Japan
关键词
electrochemical detection; detection; LC; nickel-titanium electrode; titanium-nickel electrode; electrodes; carbohydrates;
D O I
10.1016/S0021-9673(99)00047-3
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A highly sensitive amperometric detector for the liquid chromatographic analysis of carbohydrates has been developed based on the optimization of the Ni content in a Ni-Ti alloy electrode. An optimum nickel content of 30 at. % was obtained from a signal-to-noise analysis in a flow injection experiment on glucose in 0.1 M NaOH. An optimum working potential of 0.5 V vs. Ag/AgCl was obtained from a hydrodynamic voltammogram. The glucose detection limit in the flow injection analysis was 11 fmol. The electrode was applied to an LC detector with a microbore column and the detection limit was 50 fmol. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:17 / 24
页数:8
相关论文
共 11 条
[1]  
[Anonymous], **DROPPED REF**
[2]   RECENT DEVELOPMENTS IN THE CHROMATOGRAPHIC ANALYSIS OF CARBOHYDRATES [J].
CHURMS, SC .
JOURNAL OF CHROMATOGRAPHY, 1990, 500 :555-583
[3]  
GROSS AF, 1987, ANAL CHEM, V59, pR212, DOI 10.1021/ac00139a013
[4]   SIMULTANEOUS DETERMINATION OF REDUCING MONOSACCHARIDES BY CAPILLARY ZONE ELECTROPHORESIS AS THE BORATE COMPLEXES OF N-2 PYRIDYLGLYCAMINES [J].
HONDA, S ;
IWASE, S ;
MAKINO, A ;
FUJIWARA, S .
ANALYTICAL BIOCHEMISTRY, 1989, 176 (01) :72-77
[5]  
HUANG T, 1990, CURRENT SEP, V9, P139
[6]   Copper electrodes for stable subpicomole detection of carbohydrates in high-performance liquid chromatography [J].
Kano, K ;
Takagi, K ;
Inoue, K ;
Ikeda, T ;
Ueda, T .
JOURNAL OF CHROMATOGRAPHY A, 1996, 721 (01) :53-57
[7]   SENSITIVE, LASER-ASSISTED DETERMINATION OF COMPLEX OLIGOSACCHARIDE MIXTURES SEPARATED BY CAPILLARY GEL-ELECTROPHORESIS AT HIGH-RESOLUTION [J].
LIU, JP ;
SHIROTA, O ;
NOVOTNY, MV .
ANALYTICAL CHEMISTRY, 1992, 64 (08) :973-975
[8]   NICKEL-TITANIUM ALLOY ELECTRODE AS A SENSITIVE AND STABLE LC-EC DETECTOR FOR CARBOHYDRATES [J].
LUO, PFF ;
KUWANA, T .
ANALYTICAL CHEMISTRY, 1994, 66 (17) :2775-2782
[9]   ELECTROCHEMICAL DETECTION OF CARBOHYDRATES AT NICKEL-COPPER AND NICKEL-CHROMIUM-IRON ALLOY ELECTRODES [J].
MARIOLI, JM ;
KUWANA, T .
ELECTROANALYSIS, 1993, 5 (01) :11-15
[10]   Concentration of extracellular L-glutamate released from cultured nerve cells measured with a small volume online sensor [J].
Niwa, O ;
Torimitsu, K ;
Morita, M ;
Osborne, P ;
Yamamoto, K .
ANALYTICAL CHEMISTRY, 1996, 68 (11) :1865-1870