Nickel content dependence of electrochemical behavior of carbohydrates on a titanium-nickel alloy electrode and its application to a liquid chromatography detector

被引:25
作者
Morita, M
Niwa, O
Tou, S
Watanabe, N
机构
[1] NTT, Basic Res Labs, Atsugi, Kanagawa 2430198, Japan
[2] BAS Japan Ltd, Tokyo 1318790, Japan
[3] Univ Tokyo, Fac Engn, Dept Appl Chem, Tokyo 1138656, Japan
关键词
electrochemical detection; detection; LC; nickel-titanium electrode; titanium-nickel electrode; electrodes; carbohydrates;
D O I
10.1016/S0021-9673(99)00047-3
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A highly sensitive amperometric detector for the liquid chromatographic analysis of carbohydrates has been developed based on the optimization of the Ni content in a Ni-Ti alloy electrode. An optimum nickel content of 30 at. % was obtained from a signal-to-noise analysis in a flow injection experiment on glucose in 0.1 M NaOH. An optimum working potential of 0.5 V vs. Ag/AgCl was obtained from a hydrodynamic voltammogram. The glucose detection limit in the flow injection analysis was 11 fmol. The electrode was applied to an LC detector with a microbore column and the detection limit was 50 fmol. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:17 / 24
页数:8
相关论文
共 11 条
[11]   DETECTION OF CARBOHYDRATES BY CAPILLARY ELECTROPHORESIS WITH PULSED AMPEROMETRIC DETECTION [J].
OSHEA, TJ ;
LUNTE, SM ;
LACOURSE, WR .
ANALYTICAL CHEMISTRY, 1993, 65 (07) :948-951