Evaluation of single-use rotary nickel-titanium instruments

被引:90
作者
Arens, FC
Hoen, MM
Steiman, HR
Dietz, GC
机构
[1] Univ Detroit Mercy, Sch Dent, Detroit, MI 48221 USA
[2] Univ Detroit Mercy, Grad Endodont Program, Detroit, MI 48221 USA
关键词
D O I
10.1097/00004770-200310000-00013
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The purpose of this study was to analyze the number and types of defects observed in single-use, rotary nickel-titanium instruments. Every ProFile Series 29.04 taper nickel-titanium instrument used during a 4-week period in an endodontic specialty practice was collected. All instruments were new and were used by experienced clinicians during a single patient visit. The instruments were routinely used in a crown-down manner with RC Prep lubrication and copious irrigation. The instruments were used in a MicroMega 324 air motor in a 6:1 gear reduction contra-angle at 333 rpm. The instruments were collected, ultrasonically cleaned, sterilized, and inspected at x 16 magnification. Torsional, flexural, and fracture defects were recorded and statistical analysis was performed using the Kruskal-Wallis one-way analysis of variance. A total of 786 ProFile Series 29 nickel-titanium rotary instruments were evaluated; 115 (14.63%) showed some type of defect after one clinical use. Size 3 instruments had the highest defect rate (22.66%) followed by size 5 (17.30%), size 2 (17.24%), and size 4 instruments (16.10%). However, there was no statistically significant difference. The size 6 and size 7 instruments showed minimal defects (2.38% and 4.76% respectfully). Seven of 786 files had fractured (0.891%). There was no statistically significant difference in the type of failure seen within each file size. This study does show that defects can occur even with new files in the hands of experienced endodontists, and for absolute safety a single-use approach should be followed.
引用
收藏
页码:664 / 666
页数:3
相关论文
共 10 条
[1]   Comparison of nickel-titanium file distortion using electric and air-driven handpieces [J].
Bortnick, KL ;
Steiman, HR ;
Ruskin, A .
JOURNAL OF ENDODONTICS, 2001, 27 (01) :57-59
[2]   Effect of rotational speed on nickel-titanium file distortion [J].
Gabel, WP ;
Hoen, M ;
Steiman, HR ;
Pink, FE ;
Dietz, R .
JOURNAL OF ENDODONTICS, 1999, 25 (11) :752-754
[3]  
Gambarini G, 2000, ENDOD DENT TRAUMATOL, V16, P95
[4]   Cyclic fatigue of Profile rotary instruments after prolonged clinical use [J].
Gambarini, G .
INTERNATIONAL ENDODONTIC JOURNAL, 2001, 34 (05) :386-389
[5]   A reconstructed computerized tomographic comparison of Ni-Ti rotary GT™ files versus traditional instruments in canals shaped by novice operators [J].
Gluskin, AH ;
Brown, DC ;
Buchanan, LS .
INTERNATIONAL ENDODONTIC JOURNAL, 2001, 34 (06) :476-484
[6]  
Koch S.K., 2002, ORAL HLTH, V2, P39
[7]   Defects in rotary nickel-titanium files after clinical use [J].
Sattapan, B ;
Nervo, GJ ;
Palamara, JEA ;
Messer, HH .
JOURNAL OF ENDODONTICS, 2000, 26 (03) :161-165
[8]  
Thompson SA, 1997, INT ENDOD J, V30, P1, DOI 10.1111/j.1365-2591.1997.tb01093.x
[9]   Cyclic fatigue of Profile rotary instruments after clinical use [J].
Yared, GM ;
Dagher, FEB ;
Machtou, P .
INTERNATIONAL ENDODONTIC JOURNAL, 2000, 33 (03) :204-207
[10]   Cyclic fatigue of Profile rotary instruments after simulated clinical use [J].
Yared, GM ;
Dagher, FEB ;
Machtou, P .
INTERNATIONAL ENDODONTIC JOURNAL, 1999, 32 (02) :115-119