Connecting the Centrimag Levitronix pump to Berlin Heart Excor cannulae; A new approach to bridge to bridge

被引:35
作者
Maat, Alexander P. W. M. [1 ]
van Thiel, Robert J. [2 ]
Dalinghaus, Michiel [3 ,4 ]
Bogers, Ad J. J. C. [1 ]
机构
[1] Erasmus MC, Thoraxctr, Dept Cardiothorac Surg, NL-3015 CE Rotterdam, Netherlands
[2] Erasmus MC, Thoraxctr, Dept Intens Care, NL-3015 CE Rotterdam, Netherlands
[3] Erasmus MC, Thoraxctr, Dept Pediat Cardiol, NL-3015 CE Rotterdam, Netherlands
[4] Erasmus MC, Sophia Pediat Hosp, NL-3015 CE Rotterdam, Netherlands
关键词
D O I
10.1016/j.healun.2007.10.010
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: An increasing number of children are requiring circulatory suport. Hospitals offering pediatric Ventricular Assist Device (VAD) should have devices of different sizes available to cover the full range of patient sizes incurring considerable expense. As in adults, post-operative bleeding often complicates VAD implantation. The use of a Levitronix Centrimag centrifugal pump, connected to Berlin Heart Excor cannulae, seems an attractive and logic combination, both in terms of patient safety and of hospital economics. Methods: We describe 3 children with therapy resistant cardiac failure who underwent extracorporeal membrane oxygenation (ECMO) as a rescue before proceeding to placement of the Berlin Heart Excor paracorporeal assist device. The Levitronix Centrimag pump was used as an intermediate device to allow the patients to be stabilised. Therefore, only Berlin Heart cannulae of different sizes have to be readily available; if successful stabilization can be achieved, the Berlin Heart Excor ventricles and the drive unit can then be ordered to replace the Levitronix pump. Results: Two patients were successfully stabilised with the Levitronix pump and were switched to the definitive Berlin Heart Excor ventricles after 6 days of support. The third child succumbed due to intractable pulmonary hemorrhage in severely damaged lungs-No device related complications, especially no thrombo-embolic events, occurred during Levitronix support. Conclusion: The Levitronix Centrimag pump was easy to handle and gave effective circulatory support, the patients were only switched to the Berlin Heart Excor system after stabilization. In patients with a high risk of failure, it is a relatively cheap but safe and effective support system.
引用
收藏
页码:112 / 115
页数:4
相关论文
共 7 条
[1]   Pediatric bridge to heart transplantation:: Application of the Berlin heart, Medos and Thoratec ventricular assist devices [J].
Arabía, FA ;
Tsau, PH ;
Smith, RG ;
Nolan, PE ;
Paramesh, V ;
Bose, RK ;
Woolley, DS ;
Sethi, GK ;
Rhenman, BE ;
Copeland, JG .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2006, 25 (01) :16-21
[2]   Clinical performance with the levitronix centrimag short-term ventricular assist device [J].
De Robertis, F ;
Birks, EJ ;
Rogers, P ;
Dreyfus, G ;
Pepper, JR ;
Khaghani, A .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2006, 25 (02) :181-186
[3]   Circulatory support with pneumatic paracorporeal ventricular assist device in infants and children [J].
Hetzer, R ;
Loebe, M ;
Potapov, EV ;
Weng, YG ;
Stiller, B ;
Hennig, E ;
Alexi-Meskishvili, V ;
Lange, PE .
ANNALS OF THORACIC SURGERY, 1998, 66 (05) :1498-1505
[4]   Improvement in survival after mechanical circulatory support with pneumatic pulsatile ventricular assist devices in pediatric patients [J].
Hetzer, Roland ;
Potapov, Evgenij V. ;
Stiller, Brigitte ;
Weng, Yuguo ;
Huebler, Michael ;
Lemmer, Julia ;
Alexi-Meskishvili, Vladimir ;
Redlin, Matthias ;
Merkle, Frank ;
Kaufmann, Friedrich ;
Hennig, Ewald .
ANNALS OF THORACIC SURGERY, 2006, 82 (03) :917-925
[5]   Levitronix as a short-term salvage treatment for primary graft failure after heart transplantation [J].
Santise, Gianluca ;
Petrou, Mario ;
Pepper, John R. ;
Dreyfus, Gilles ;
Khaghani, Asghar ;
Birks, Emma J. .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2006, 25 (05) :495-498
[6]   Management of pediatric patients after implantation of the Berlin Heart EXCOR ventricular assist device [J].
Stiller, Brigitte ;
Lemmer, Julia ;
Schubert, Stephan ;
Ewert, Peter ;
Schulze-Neick, Ingram ;
Huebler, Michael ;
Redlin, Matthias ;
Berger, Felix ;
Hetzer, Roland .
ASAIO JOURNAL, 2006, 52 (05) :497-500
[7]  
2006, ISHLT MONOGRAPH SERI, V1, P127