Impact of membrane desolvation on the effects of sodium on response in inductively coupled plasma atomic emission spectrometry with ultrasonic sample introduction

被引:8
作者
Allen, LB [1 ]
Siitonen, PH [1 ]
Thompson, HC [1 ]
机构
[1] US FDA, Natl Ctr Toxicol Res, Div Chem, Jefferson, AR 72079 USA
关键词
inductively coupled plasma atomic emission spectrometry; ultrasonic sample introduction; microporous membrane desolvation; easily ionizable elements;
D O I
10.1039/a800837j
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This report compares the influence of sodium on response with ICP-AES for copper, cadmium, lead and cobalt with ultrasonic sample introduction in the presence and absence of a microporous membrane desolvation (MMD). Comparisons are reported at viewing positions of 8, 14, and 20 mm above the load coil, for injector gas flow rates of 0, 70, 0.85 and 1.00 l min(-1), and at sodium concentrations ranging from 0.00 to 0.08 M Na. Blank emission intensities were depressed in the presence of sodium regardless of the operating condition(s), Depressions in the blank emission intensities were greater without the MMD, but were less than 20%, The magnitude of depression in the blank emission intensities did not vary significantly among the analyte lines or with the viewing position, The impact of sodium on signal intensities with and without the MMD was greater than the impact on the blank emission intensities, with changes approaching 150%, At the lower viewing positions, signal enhancement in the presence of sodium was common and the degree of enhancement was greater with the MMD, Response changes with respect to sodium concentration were also different with and without the MMD low in the plasma. Higher in the plasma, element-to-element variation occurred in the nature of the impact of sodium, For signals that were enhanced, the degree of enhancement was generally greater with the MMD, but when signals were depressed, there was no correlation to the presence or absence of the MMD. In contrast to the lower viewing position, higher in the plasma response changes with respect to the sodium concentration were similar. These results suggest that although the plasma is physically different with and without the MMD, the MMD does not alter the mechanism(s) by which sodium impacts response.
引用
收藏
页码:735 / 741
页数:7
相关论文
共 18 条
[1]   Aerosol-phase assisted sample digestion for the determination of trace metals in organic samples by plasma atomic emission spectrometry [J].
Allen, LB ;
Siitonen, PH ;
Thompson, HC .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (07) :529-532
[2]  
Blades M. W., 1994, SPECTROSCOPY, V9, P14
[3]   INTERFERENCE FROM EASILY IONIZABLE ELEMENT MATRICES IN INDUCTIVELY COUPLED PLASMA EMISSION-SPECTROMETRY - A SPATIAL STUDY [J].
BLADES, MW ;
HORLICK, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1981, 36 (09) :881-900
[4]   Universal calibration for analysis of organic solutions by inductively coupled plasma atomic emission spectrometry [J].
Botto, RI ;
Zhu, JJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (09) :675-681
[5]   USE OF AN ULTRASONIC NEBULIZER WITH MEMBRANE DESOLVATION FOR ANALYSIS OF VOLATILE SOLVENTS BY INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROMETRY [J].
BOTTO, RI ;
ZHU, JJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (09) :905-912
[6]  
Brenner IB, 1996, FRESEN J ANAL CHEM, V355, P774
[7]   PERFORMANCE-CHARACTERISTICS OF AN ULTRASONIC NEBULIZER COUPLED TO A 40.68 MHZ INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION SPECTROMETER [J].
BRENNER, IB ;
BREMIER, P ;
LEMARCHAND, A .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1992, 7 (06) :819-824
[8]   EASILY IONIZABLE ELEMENT INTERFERENCES IN INDUCTIVELY-COUPLED PLASMA-ATOMIC EMISSION-SPECTROSCOPY .1. EFFECT ON RADIAL ANALYTE EMISSION PATTERNS [J].
GALLEY, PJ ;
GLICK, M ;
HIEFTJE, GM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1993, 48 (6-7) :769-788
[9]   ULTRASONIC NEBULIZATION IN INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
GALLI, P ;
ODDO, N .
MICROCHEMICAL JOURNAL, 1992, 46 (03) :327-334
[10]   THE EFFECT OF DESOLVATING DROPLETS AND VAPORIZING PARTICLES ON IONIZATION AND EXCITATION IN AR INDUCTIVELY-COUPLED PLASMAS [J].
HOBBS, SE ;
OLESIK, JW .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1993, 48 (6-7) :817-833