Modelling motor cortex stimulation for chronic pain control: electrical potential field, activating functions and responses of simple nerve fibre models

被引:91
作者
Manola, L
Roelofsen, BH
Holsheimer, J [1 ]
Marani, E
Geelen, J
机构
[1] Univ Twente, Inst Biomed Technol, NL-7500 AE Enschede, Netherlands
[2] Leiden Univ, Med Ctr, Dept Neurosurg, Neuroregulat Grp, Leiden, Netherlands
[3] Med Spectrum Twente Hosp, Dept Neurol, Enschede, Netherlands
关键词
motor cortex stimulation; chronic pain; volume conductor model; nerve fibre model; activating function;
D O I
10.1007/BF02345810
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
This computer modelling study on motor cortex stimulation (MCS) introduced a motor cortex model, developed to calculate the imposed electrical potential field characteristics and the initial response of simple fibre models to stimulation of the precentral gyrus by an epidural electrode, as applied in the treatment of chronic, intractable pain. The model consisted of two parts: a three-dimensional volume conductor based on tissue conductivities and human anatomical data, in which the stimulation-induced potential field was computed, and myelinated nerve fibre models allowing the calculation of their response to this field. A simple afferent fibre branch and three simple efferent fibres leaving the cortex at different positions in the precentral gyrus were implemented. It was shown that the thickness of the cerebrospinal fluid (CSF) layer between the dura mater and the cortex below the stimulating electrode substantially affected the distribution of the electrical potential field in the precentral gyrus and thus the threshold stimulus for motor responses and the therapeutic stimulation amplitude. When the CSF thickness was increased from 0 to 2.5 mm, the load impedance decreased by 28%, and the stimulation amplitude increased by 6.6 V for each millimetre of CSF. Owing to the large anode-cathode distance (10 mm centre-to-centre) in MCS, the cathodal fields in mono- and bipolar stimulation were almost identical. Calculation of activating functions and fibre responses showed that only nerve fibres with a directional component parallel to the electrode surface were excitable by a cathode, whereas fibres perpendicular to the electrode surface were excitable under an anode.
引用
收藏
页码:335 / 343
页数:9
相关论文
共 31 条
[21]   ANALYSIS OF MODELS FOR EXTERNAL STIMULATION OF AXONS [J].
RATTAY, F .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1986, 33 (10) :974-977
[22]  
SMITH GD, 2001, THALAMUS RELAT SYST, V11, P1
[23]   RECRUITMENT OF DORSAL COLUMN FIBERS IN SPINAL-CORD STIMULATION - INFLUENCE OF COLLATERAL BRANCHING [J].
STRUIJK, JJ ;
HOLSHEIMER, J ;
VANDERHEIDE, GG ;
BOOM, HBK .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1992, 39 (09) :903-912
[24]   A stereological method for estimating the total length and size of myelin fibers in human brain white matter [J].
Tang, Y ;
Nyengaard, JR .
JOURNAL OF NEUROSCIENCE METHODS, 1997, 73 (02) :193-200
[25]  
TRUEX RC, 1964, HUMAN ANATOMY
[26]   CHRONIC MOTOR CORTEX STIMULATION IN PATIENTS WITH THALAMIC PAIN [J].
TSUBOKAWA, T ;
KATAYAMA, Y ;
YAMAMOTO, T ;
HIRAYAMA, T ;
KOYAMA, S .
JOURNAL OF NEUROSURGERY, 1993, 78 (03) :393-401
[27]  
VILLIGER L, 1940, GEHIRN RUCKENMARK
[28]  
VONKEYSERLINGK DG, 1984, ANAT ANZEIGER, V157, P97
[29]   A model of the electrical behaviour of myelinated sensory nerve fibres based on human data [J].
Wasselink, WA ;
Holsheimer, J ;
Boom, HBK .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1999, 37 (02) :228-235
[30]  
ZEFFIRO TA, 1990, HUMAN NERVOUS SYSTEM, P803