Conversion of a sequential inductively coupled plasma emission spectrometer into a multichannel simultaneous system using a photodiode array detector

被引:5
作者
Pimentel, MF
de Araujo, MCU
de Barros Neto, B
Pasquini, C
机构
[1] ITEP, Fundacao Inst Tecnol Estado Pernambuco, BR-50740540 Recife, PE, Brazil
[2] Univ Fed Paraiba, CCEN, Dept Quim, BR-58059900 Joao Pessoa, Paraiba, Brazil
[3] Univ Fed Pernambuco, CCEN, Dept Quim Fundamental, BR-50739907 Recife, PE, Brazil
[4] Univ Estadual Campinas, UNICAMP, Inst Quim, BR-13081970 Campinas, SP, Brazil
来源
JOURNAL OF AUTOMATIC CHEMISTRY | 1998年 / 20卷 / 03期
关键词
D O I
10.1155/S146392469800008X
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A monochannel plasma emission spectrometer was converted to a multichannel instrument by the introduction of a detection system based on an array of 1024 photodiodes and a low-resolution dispersion device. The new, relatively inexpensive equipment, features both the high speed typical of simultaneous instruments and the versatility of scanning systems. This paper reports on an evaluation of the modified equipment for quantitative analysis with the simultaneous determination of Al, Mn, Mg, Ca, Fe and Cu in a natural water matrix. An average relative prediction error of 2.4% was found which is the same as the error obtained with the conventional analytical method. Data acquisition with the modified instrument is up to 40 times faster.
引用
收藏
页码:69 / 75
页数:7
相关论文
共 35 条
[1]   SOLID-STATE DETECTOR FOR ICP-OES [J].
BARNARD, TW ;
CROCKETT, MI ;
IVALDI, JC ;
LUNDBERG, PL ;
YATES, DA ;
LEVINE, PA ;
SAUER, DJ .
ANALYTICAL CHEMISTRY, 1993, 65 (09) :1231-1239
[2]   DESIGN AND EVALUATION OF AN ECHELLE GRATING OPTICAL-SYSTEM FOR ICP-OES [J].
BARNARD, TW ;
CROCKETT, MI ;
IVALDI, JC ;
LUNDBERG, PL .
ANALYTICAL CHEMISTRY, 1993, 65 (09) :1225-1230
[3]  
BELATTO CR, 1996, J AUTOM CHEM, V18, P7
[4]   ELEMENTAL ANALYSIS WITH A PLASMA EMISSION ECHELLE SPECTROMETER EMPLOYING A CHARGE INJECTION DEVICE (CID) DETECTOR [J].
BILHORN, RB ;
DENTON, MB .
APPLIED SPECTROSCOPY, 1989, 43 (01) :1-11
[5]   SPECTROCHEMICAL MEASUREMENTS WITH MULTICHANNEL INTEGRATING DETECTORS [J].
BILHORN, RB ;
EPPERSON, PM ;
SWEEDLER, JV ;
DENTON, MB .
APPLIED SPECTROSCOPY, 1987, 41 (07) :1125-1136
[6]   APPLICATION OF A PHOTODIODE ARRAY DETECTOR TO AN ARC SPECTROGRAPH [J].
BRETT, L ;
STAHL, RG ;
TIMMINS, KJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (04) :333-336
[7]  
BRUSHWYLER KR, 1990, TALANTA, V37, P23
[8]   MULTIPLE ENTRANCE SLIT VIDICON SPECTROMETER FOR SIMULTANEOUS MULTIELEMENT ANALYSIS [J].
BUSCH, KW ;
MALLOY, B ;
TALMI, Y .
ANALYTICAL CHEMISTRY, 1979, 51 (06) :670-673
[9]  
BUSCH KW, 1990, MULTIELEMENT DETECTI
[10]  
Draper N. R., 1966, APPL REGRESSION ANAL