Cochlear structures after implantation of a perimodiolar electrode array

被引:31
作者
Richter, B [1 ]
Jaekel, K [1 ]
Aschendorff, A [1 ]
Marangos, N [1 ]
Laszig, R [1 ]
机构
[1] Univ Clin Freiburg, Dept Otolaryngol Head & Neck Surg, D-79106 Freiburg, Germany
关键词
perimodiolar CI-electrode array; human temporal bones; intracochlear damage; Teflon (TM) positioner; explantation; re-implantation;
D O I
10.1097/00005537-200105000-00015
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Objective: A perimodiolar prototype electrode array has been evaluated on 12 human temporal bone specimens, To achieve modiolar proximity, a fine Tenon(TM) positioner is attached to the tip and base of a conventional Nucleus 22 electrode array. Methods: Intracochlear electrode position was examined by macro scopic preparation and confirmed by a cochlear view x-ray after insertion into human temporal bones. The temporal bones were examined with light microscopy for intracochlear trauma after insertion, explantation, and re-implantation of the perimodiolar array. Results: Macroscopic preparation and x-ray confirmed close proximity of the intracochlear electrodes to the modiolar wall. In some bones, electrode insertion and positioning was performed without significant damage to intracochlear structures, but explantation caused varying degrees of trauma depending on the explantation technique used. Re-implantation and repeat explantation appeared to destroy most of the intracochlear architecture. Conclusions: The perimodiolar electrode array prototype with Tenon positioner cannot be recommended for clinical use, especially in children in whom the possibility for re-implantation is high.
引用
收藏
页码:837 / 843
页数:7
相关论文
共 22 条
[11]  
KENNEDY DW, 1987, LARYNGOSCOPE, V97, P42
[12]   COCHLEAR REIMPLANTATION [J].
SAEED, SR ;
RAMSDEN, RT ;
HARTLEY, C ;
WOOLFORD, TJ ;
BOYD, P .
JOURNAL OF LARYNGOLOGY AND OTOLOGY, 1995, 109 (10) :980-985
[13]   CHRONIC INTRACOCHLEAR ELECTRODE IMPLANTATION - COCHLEAR PATHOLOGY AND ACOUSTIC NERVE SURVIVAL [J].
SCHINDLER, RA ;
MERZENICH, MM .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1974, 83 (02) :202-215
[14]  
SCHINDLER RA, 1977, ARCH OTOLARYNGOL, V103, P691
[15]  
SCHINDLER RA, 1975, J LARYNGOL OTOL, V90, P445
[16]  
SCHUKNECHT HF, 1953, AMA ARCH OTOLARYNGOL, V58, P377
[17]   BANDED INTRACOCHLEAR ELECTRODE ARRAY - EVALUATION OF INSERTION TRAUMA IN HUMAN TEMPORAL BONES [J].
SHEPHERD, RK ;
PYMAN, BC ;
CLARK, GM ;
WEBB, RL .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1985, 94 (01) :55-59
[18]   ELECTRICAL-STIMULATION OF THE AUDITORY-NERVE - THE EFFECT OF ELECTRODE POSITION ON NEURAL EXCITATION [J].
SHEPHERD, RK ;
HATSUSHIKA, S ;
CLARK, GM .
HEARING RESEARCH, 1993, 66 (01) :108-120
[19]   ANATOMICAL AND PATHOLOGICAL ASPECTS OF THE ELECTRICAL-STIMULATION OF THE DEAF INNER-EAR [J].
SPOENDLIN, H .
ARCHIVES OF OTO-RHINO-LARYNGOLOGY-ARCHIV FUR OHREN-NASEN-UND KEHLKOPFHEILKUNDE, 1979, 223 (01) :1-75
[20]   Comparison of electrode position in the human cochlea using various perimodiolar electrode arrays [J].
Tykocinski, M ;
Cohen, LT ;
Pyman, BC ;
Roland, T ;
Treaba, C ;
Palamara, J ;
Dahm, MC ;
Shepherd, RK ;
Xu, J ;
Cowan, RS ;
Cohen, NL ;
Clark, GM .
AMERICAN JOURNAL OF OTOLOGY, 2000, 21 (02) :205-211