Electrical conductivity imaging by magnetic resonance electrical impedance tomography (MREIT)

被引:19
作者
Oh, SH
Han, JY
Lee, SY
Cho, MH
Lee, BI
Woo, EJ
机构
[1] Kyung Hee Univ, Grad Sch East West Med Sci, Kyungki 449701, South Korea
[2] Kyung Hee Univ, Coll Elect & Informat, Kyungki, South Korea
关键词
electrical conductivity imaging; current density imaging; MRCDI; MREIT; J-substitution algorithm;
D O I
10.1002/mrm.10588
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Magnetic resonance electrical impedance tomography (MREIT) is a recently developed imaging technique that combines MRI and electrical impedance tomography (EIT). In MREIT, cross-sectional electrical conductivity images are reconstructed from the internal magnetic field density data produced inside an electrically conducting subject when an electrical current is injected into the subject. In this work the results of an electrical conductivity imaging experiment are presented, along with some practical considerations regarding MREIT. The MREIT experiment was performed with a 0.3 Tesla MRI system on a phantom made of two compartments with different electrical conductivities. The current density inside the phantom was measured by the MR current density imaging (MRCDI) technique. The measured current density was then used for conductivity image reconstruction by the J-substitution algorithm. The conductivity phantom images obtained with an injection current of 28mA showed conductivity errors of about 25.5%. (C) 2003 Wiley-Liss, Inc.
引用
收藏
页码:875 / 878
页数:4
相关论文
共 17 条
[1]   Use of a priori information in estimating tissue resistivities-application to measured data [J].
Baysal, U ;
Eyüboglu, BM .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (07) :1677-1689
[2]  
BAYSAL U, 2002, P 13 INT C BIOM JEN, P682
[3]  
BIRGUL O, 2001, P 23 ANN C IEEE EMBS
[4]  
DEMONTE TP, 2002, P 10 SCI M ISMRM HON, P2313
[5]   Measurement of electrical current density distribution within the tissues of the head by magnetic resonance imaging [J].
Gamba, HR ;
Delpy, DT .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1998, 36 (02) :165-170
[6]  
Ider YZ., 1998, ELEKTRIK, V6, P215
[7]   J-substitution algorithm in Magnetic Resonance Electrical Impedance Tomography (MREIT):: Phantom experiments for static resistivity images [J].
Khang, HS ;
Lee, BI ;
Oh, SH ;
Woo, EJ ;
Lee, SY ;
Cho, MY ;
Kwon, O ;
Yoon, JR ;
Seo, JK .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2002, 21 (06) :695-702
[8]  
Khang HS, 2002, MEAS SCI TECHNOL, V13, pN42, DOI 10.1088/0957-0233/13/4/403
[9]   Magnetic resonance electrical impedance tomography (MREIT):: Simulation study of J-substitution algorithm [J].
Kwon, O ;
Woo, EJ ;
Yoon, JR ;
Seo, JK .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2002, 49 (02) :160-167
[10]  
OH SH, 2002, P 10 SCI M ISMRM HON, P2315