Speckle reduction and contrast enhancement of echocardiograms via multiscale nonlinear processing

被引:219
作者
Zong, XL
Laine, AF
Geiser, EA
机构
[1] Univ Florida, Dept Comp Sci & Informat Engn, Gainesville, FL 32611 USA
[2] Columbia Univ, Ctr Biomed Engn, New York, NY 10027 USA
[3] Univ Florida, Dept Med, Echocardiog Res Lab, Gainesville, FL 32610 USA
关键词
contrast enhancement; denoising; echocardiograms; multiscale representations; nonlinear processing; speckle reduction; ultrasound images; wavelet shrinkage;
D O I
10.1109/42.730398
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
This paper presents an algorithm for speckle reduction and contrast enhancement of echocardiographic images. Within a framework of multiscale wavelet analysis, we apply wavelet shrinkage techniques to eliminate: noise while preserving the sharpness of salient features. In addition, nonlinear processing of feature energy is carried out to enhance contrast within local structures and along object boundaries. We show that the algorithm is capable of not only reducing speckle, but also enhancing features of diagnostic importance, such as myocardial walls in two-dimensional echocardiograms obtained from the parasternal short-axis view. Shrinkage of wavelet coefficients via soft thresholding: within finer levels of scale is carried out on coefficients of logarithmically transformed echocardiograms, Enhancement of echocardiographic features is accomplished via nonlinear stretching followed by hard thresholding of wavelet coefficients within selected (midrange) spatial-frequency levels of analysis. We formulate the denoising and enhancement problem, introduce a class of dyadic wavelets, and describe our implementation of a dyadic wavelet transform. Our approach for speckle reduction and contrast enhancement was shown to be less affected by pseudo-Gibbs phenomena, We show experimentally that this technique produced superior results both qualitatively and quantitatively when compared to results obtained from existing denoising methods alone. A study using a database of clinical echocardiographic images suggests that such denoising and enhancement may improve the overall consistency of expert observers to manually defined borders.
引用
收藏
页码:532 / 540
页数:9
相关论文
共 30 条
[1]   ACOUSTIC SPECKLE - THEORY AND EXPERIMENTAL-ANALYSIS [J].
ABBOTT, JG ;
THURSTONE, FL .
ULTRASONIC IMAGING, 1979, 1 (04) :303-324
[2]   A multiple active contour model for cardiac boundary detection on echocardiographic sequences [J].
Chalana, V ;
Linker, DT ;
Haynor, DR ;
Kim, YM .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1996, 15 (03) :290-298
[3]  
COIFMAN R. R., 1995, Wavelets and statistics, P125, DOI DOI 10.1007/978-1-4612-2544-7_9
[4]  
COIFMAN RR, 1993, PROGRESS IN WAVELET ANALYSIS AND APPLICATIONS, P63
[5]   COMPUTER-ASSISTED EDGE-DETECTION IN TWO-DIMENSIONAL ECHOCARDIOGRAPHY - COMPARISON WITH ANATOMIC DATA [J].
COLLINS, SM ;
SKORTON, DJ ;
GEISER, EA ;
NICHOLS, JA ;
CONETTA, DA ;
PANDIAN, NG ;
KERBER, RE .
AMERICAN JOURNAL OF CARDIOLOGY, 1984, 53 (09) :1380-1387
[6]   ORTHONORMAL BASES OF COMPACTLY SUPPORTED WAVELETS [J].
DAUBECHIES, I .
COMMUNICATIONS ON PURE AND APPLIED MATHEMATICS, 1988, 41 (07) :909-996
[7]   DE-NOISING BY SOFT-THRESHOLDING [J].
DONOHO, DL .
IEEE TRANSACTIONS ON INFORMATION THEORY, 1995, 41 (03) :613-627
[8]  
DONOHO DL, 1992, 400 STAN U DEP STAT
[9]  
FAN J, 1996, WAVELETS MED BIOL, P163
[10]  
Geiser E A, 1991, J Am Soc Echocardiogr, V4, P235