Current-based transthoracic defibrillation

被引:20
作者
Kerber, RE
Kieso, RA
Kienzle, MG
Olshansky, B
Waldo, AL
Carlson, MD
Wilber, DJ
Aschoff, AM
Birger, S
Charbonnier, F
机构
[1] LOYOLA UNIV,CHICAGO,IL 60611
[2] CASE WESTERN RESERVE UNIV HOSP,CLEVELAND,OH
[3] HEWLETT PACKARD CORP,MCMINNVILLE,OR
关键词
D O I
10.1016/S0002-9149(96)90062-4
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
This study examines in a prospective, multicenter trial the feasibility and advantage of current-based, transthoracic defibrillation. Current-based, damped, sinusoidal waveform shocks of 18, 25, 30, 35, or 40 amperes (A) were administered beginning with 25 A for polymorphic ventricular tachycardia (VT) and ventricular fibrillation (VF) or 18 A for monomorphic VT; success rates were compared with those of energy-based shocks beginning at 200 J for VF/polymorphic VT and 100 J for VT. The current-based shocks were delivered from custom-modified defibrillators that determined impedance in advance of any shock using a ''test-pulse'' technique; the capacitor then charged to the exact energy necessary to deliver the operator-selected current against the impedance determined by the defibrillator. Three hundred sixty-two patients received >1 shock for VF, polymorphic VT, or monomorphic VT: 569 current- based shocks and 420 energy-based shocks. Current-based shocks of 35/40 A achieved success rates of up to 74% for VF/polymorphic VT; 30 A shocks terminated 88% of monomorphic VT episodes. Energy-based shocks of 300 J terminated 72% of VF/polymorphic VT; 200-J shocks terminated 89% of monomorphic VT. We could not demonstrate a significant increase in the success rate of current-based shocks over energy-based shocks for patients with high transthoracic impedance; this may be due to inadequate sample size. Thus, current-based defibrillation is clinically feasible and effective. A larger study will be needed to test whether current-based defibrillation is superior to energy-based defibrillation. (C) 1996 by Excerpta Medica, Inc.
引用
收藏
页码:1113 / 1118
页数:6
相关论文
共 17 条
[1]  
DALZELL GWN, 1989, BRIT HEART J, V61, P502
[2]   MYOCARDIAL UPTAKE OF TECHNETIUM-99M STANNOUS PYROPHOSPHATE FOLLOWING DIRECT-CURRENT TRANSTHORACIC COUNTERSHOCK [J].
DICOLA, VC ;
FREEDMAN, GS ;
DOWNING, SE ;
ZARET, BL .
CIRCULATION, 1976, 54 (06) :980-986
[3]  
EWY EA, 1974, CIRCULATION S2, V49, P24
[4]  
FLEISS JL, 1981, STATISTICAL METHODS, P138
[5]  
Geddes L A, 1976, Med Instrum, V10, P159
[6]   COMPARISON OF MONOPHASIC AND BIPHASIC DEFIBRILLATING PULSE WAVE-FORMS FOR TRANSTHORACIC CARDIOVERSION [J].
GREENE, HL ;
DIMARCO, JP ;
KUDENCHUK, PJ ;
SCHEINMAN, MM ;
TANG, ASL ;
REITER, MJ ;
ECHT, DS ;
CHAPMAN, PD ;
JAZAYERI, MR ;
CHAPMAN, FW ;
AHMED, M ;
JOHNSON, JL ;
NISKANEN, RA .
AMERICAN JOURNAL OF CARDIOLOGY, 1995, 75 (16) :1135-1139
[7]   TRANS-THORACIC RESISTANCE IN HUMAN DEFIBRILLATION - INFLUENCE OF BODY-WEIGHT, CHEST SIZE, SERIAL SHOCKS, PADDLE SIZE AND PADDLE CONTACT PRESSURE [J].
KERBER, RE ;
GRAYZEL, J ;
HOYT, R ;
MARCUS, M ;
KENNEDY, J .
CIRCULATION, 1981, 63 (03) :676-682
[8]   ENERGY, CURRENT, AND SUCCESS IN DEFIBRILLATION AND CARDIOVERSION - CLINICAL-STUDIES USING AN AUTOMATED IMPEDANCE-BASED METHOD OF ENERGY ADJUSTMENT [J].
KERBER, RE ;
MARTINS, JB ;
KIENZLE, MG ;
CONSTANTIN, L ;
OLSHANSKY, B ;
HOPSON, R ;
CHARBONNIER, F .
CIRCULATION, 1988, 77 (05) :1038-1046
[9]   ADVANCE PREDICTION OF TRANS-THORACIC IMPEDANCE IN HUMAN DEFIBRILLATION AND CARDIOVERSION - IMPORTANCE OF IMPEDANCE IN DETERMINING THE SUCCESS OF LOW-ENERGY SHOCKS [J].
KERBER, RE ;
KOUBA, C ;
MARTINS, J ;
KELLY, K ;
LOW, R ;
HOYT, R ;
FERGUSON, D ;
BAILEY, L ;
BENNETT, P ;
CHARBONNIER, F .
CIRCULATION, 1984, 70 (02) :303-308
[10]   VENTRICULAR-TACHYCARDIA RATE AND MORPHOLOGY DETERMINE ENERGY AND CURRENT REQUIREMENTS FOR TRANSTHORACIC CARDIOVERSION [J].
KERBER, RE ;
KIENZLE, MG ;
OLSHANSKY, B ;
WALDO, AL ;
WILBER, D ;
CARLSON, MD ;
ASCHOFF, AM ;
BIRGER, S ;
FUGATT, L ;
WALSH, S ;
ROCKWELL, M ;
CHARBONNIER, F .
CIRCULATION, 1992, 85 (01) :158-163